Power Express, REF B90918
Recall Z-2978-2024, part of FDA recall event 95064. See Beckman Coulter Biomedical's company page.
Beckman Coulter Biomedical GmbH started this Class II device recall on Jul 17, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jul 17, 2024
- Status: Ongoing
- Products in this event: 3
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
- Product
- Power Express, REF B90918
- Quantity
- —
- Lots and codes
- UDI: (01)15099590681555/ All serial numbers
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United Kingdom, Vietnam.
- Recalling firm
- Beckman Coulter Biomedical GmbH, Munchen, Germany
- How the firm notified
- Letter
- FDA classified
- Sep 5, 2024
- PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.Z-2980-2024 · Class II · Jul 17, 2024 · Ongoing
- PE Centrifuge, REF: B36365, a component of the Power ExpressZ-2979-2024 · Class II · Jul 17, 2024 · Ongoing
- REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager…Z-2834-2026 · Class II · Apr 29, 2026 · Ongoing
- Beckman Coulter DxA Automation System, DxA 5000, REF: B87352, IVD, 200240V, 16A, 50/60HZ, RxOnlyZ-2144-2023 · Class II · Jun 2, 2023 · Ongoing
- DxA 5000; Catalog Number:B50516; UDI:15099590707118 - Product Usage: The DxA 5000 is a high-speed, modular, automated…Z-2983-2020 · Class II · Jun 17, 2020 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports