Medtronic, Evera Implantable Cardioverter Defibrillators: Product Model EVERA XT DDBB1D1
Recall Z-2651-2016, part of FDA recall event 74923. See Medtronic, Cardiac Rhythm and Heart Failure's company page.
Medtronic Inc., Cardiac Rhythm and Heart Failure started this Class II device recall on Aug 12, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 20, 2017.
- Classification: Class II
- Started: Aug 12, 2016
- Status: Terminated
- Products in this event: 2
78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.
- Product
- Medtronic, Evera Implantable Cardioverter Defibrillators: Product Model EVERA XT DDBB1D1, DDBB1D4, DDBB2D1, DDBB2D4, DVBB2D4, DVBB1D1. EVERA S DDBC3D1, DDBC3D4 Product Usage: The Medtronic Evera DR dual chamber and Evera VR single chamber, implantable cardioverter defibrillator (lCD) are a family of multiprogrammable cardiac devices that monitor and regulate the patients heart rate by providing dual or single chamber rate-responsive bradycardia pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies (dual-chamber devices only).
- Quantity
- 78 (39 US, 39 OUS)
- Lots and codes
- Product Name: Evera XT DR lCD. Models: DDBB1D4, DDBB1D1, DDBB2D4, DDBB2D1. Serial Numbers: BWC202738H, BWC202754H, BWB202998H, BWB203157H, BWB203167H, BWB203173H, BWB203186H, BWE601558S, BWE601571S, BWE601578S, BWE601579S, BWE601581S, BWE601589S, BWE601591S, BWE601594S, BWE601600S, BWE601605S, BWD602122S. Product Name: Evera S DR lCD. Models: DDBC3D1, DDBC3D4. Serial Numbers: BWG600597S, BWF600969S, BWF600970S, BWF600972S, BWF600973S, BWF600975S, BWF600977S, BWF600978S, BWF600979S, BWF600983S, BWF600984S, BWF600985S, BWF600987S, BWF600989S, BWF600991S, BWF600992S, BWF600996S, BWF601001S.…
- Distribution
- Worldwide Distribution - US Nationwide in the states of US: AZ, CA, FL, GA, IL, KY, MD, MN, NJ, NY, NC, OH, OK, PA, PR, TX. and countries of: Australia, Finland, France, Germany, Iran, Italy, Japan, Korea, Netherlands, Puerto Rico, Saudi Arabia, Spain, and United Kingdom.
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure, Mounds View, MN, United States
- How the firm notified
- Letter
- FDA classified
- Aug 24, 2016
- Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators: Product Model VIVAXTCRT-D …Z-2650-2016 · Class II · Aug 12, 2016 · Terminated
- MyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The…Z-0399-2018 · Class II · Jul 14, 2017 · Terminated
- Amplia MRI Quad CRT-0 SureScan Claria MRI Quad CRT-D SureScan Compia MRI Quad CRT-D SureScanZ-2117-2017 · Class II · Apr 12, 2017 · Terminated
- Amplia MRI" CRT -D SureScan", Amplia MRI" Quad CRT -D SureScan", Claria MRI" CRT -D SureScan", Claria MRI" Quad…Z-0962-2017 · Class II · Dec 9, 2016 · Terminated
- Medtronic, Temporary Pacing Lead System, Model 6416, Sterile EO. Product Catalog Numbers: Five (5) Pack Single Pack…Z-2200-2016 · Class II · Jun 16, 2016 · Terminated
- Medtronic, MyCareLink Patient Monitor, Model 24950, Rx Only. The MyCareLink Monitor, Model 24950, is a remote…Z-2125-2016 · Class II · May 26, 2016 · Terminated
- Medtronic CareLink" Monitor (2490C) and Medtronic CareLink Express" Monitor (2020B) Product Usage: The 2490C…Z-1605-2016 · Class II · Mar 31, 2016 · Terminated