Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators: Product Model…

Recall Z-2650-2016, part of FDA recall event 74923. See Medtronic, Cardiac Rhythm and Heart Failure's company page.

Medtronic Inc., Cardiac Rhythm and Heart Failure started this Class II device recall on Aug 12, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 20, 2017.

  • Classification: Class II
  • Started: Aug 12, 2016
  • Status: Terminated
  • Products in this event: 2

Why It Was Recalled

Class II

78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.

Record

Product
Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators: Product Model VIVAXTCRT-D DTBA1D1, DTBA1D4 VIVASCRT-D DTBB1D1, DTBB1D4 Product Usage: The Medtronic Viva CRT -D dual chamber, implantable cardioverter defibrillator with cardiac resynchronization therapy (CRT -D) is a multiprogrammable cardiac device that monitors and regulates the patients heart rate by providing single or dual chamber, rate responsive bradycardia pacing; sequential biventricular pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies.
Quantity
78 (39 US, 39 OUS)
Lots and codes
Product Name: Viva XT CRT -D . Models: DTBA1D4, DTBA1D1. Serial Numbers: BLF203128H, BLF204746H, BLE202888H, BLE202889H, BLE202890H, BLE202901H, BLE202941H, BLE202947H, BLE202954H, BLE202958H, BLE202961H, BLE202962H, BLE202964H, BLE202981H, BLE202987H, BLE202989H, BLE202990H, BLE202991H, BLE203019H, BLE203026H, BLE203027H, BLE203029H, BLE203032H, BLE203046H, BLE203052H, BLE203073H Product Name: Viva S CRT-D. Models: DTBB1D4, DTBB1D1. Serial Numbers: BLO202272H, BLN202206H.
Distribution
Worldwide Distribution - US Nationwide in the states of US: AZ, CA, FL, GA, IL, KY, MD, MN, NJ, NY, NC, OH, OK, PA, PR, TX. and countries of: Australia, Finland, France, Germany, Iran, Italy, Japan, Korea, Netherlands, Puerto Rico, Saudi Arabia, Spain, and United Kingdom.
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure, Mounds View, MN, United States
How the firm notified
Letter
FDA classified
Aug 24, 2016

Other Products in This Recall

Recall event 74923

Medtronic Inc., Cardiac Rhythm and Heart Failure's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports