Medtronic, Temporary Pacing Lead System
Recall Z-2200-2016, part of FDA recall event 74424. See Medtronic, Cardiac Rhythm and Heart Failure's company page.
Medtronic Inc., Cardiac Rhythm and Heart Failure started this Class II device recall on Jun 16, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Feb 13, 2018.
- Classification: Class II
- Started: Jun 16, 2016
- Status: Terminated
This Pacing Lead System does not comply with Section 8.5.2.3 of IEC 60601-1, and corresponding provisions of FDA 21 CFR 898, which relate to design standards to prevent connecting a patients lead to a possible hazardous voltage.
- Product
- Medtronic, Temporary Pacing Lead System, Model 6416, Sterile EO. Product Catalog Numbers: Five (5) Pack Single Pack 6416-100 6416-1005 6416-140 6416-1405 6416-200 6416-2005 For temporary intracardiac pacing and EGM recording. The system is disposable, for temporary single patient use, with a contemplated implant duration of 7 days or less. The system is used with Medtronic temporary external pacemakers.
- Quantity
- 11697 (5523 OUS, 6174 US)
- Lots and codes
- Model 6416 Lot Number List: All Lots manufactured after 01May2014 as of 23May2016. Expiration dates through 12 April 2018. AAG102212F AAG127214F AAG154212F AAG177226F AHH123209F AHH167209F AAG102219F AAG128219F AAG154219F AAG177229F AHH124209F AHH168208F AAG103213F AAG129214F AAG155210F AAG178227F AHH125207F AHH169209F AAG104212F AAG130210F AAG156210F AAG178230F AHH126209F AHH170208F AAG104219F AAG130213F AAG156213F AAG179228F AHH127209F AHH171209F AAG105210F AAG131212F AAG157212F AAG179231F AHH128209F AHH172207F AAG105213F AAG131219F AAG157219F AAG180227F AHH129209F AHH173208F AAG106219F…
- Distribution
- Worldwide Distribution. US nationwide, Albania, Algeria, Australia, Botswana, Bulgaria, Canada, Colombia, France, Gabon, Germany, Guam, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Kazakhstan, Macedonia, Mauritius, Netherlands, Netherlands Antilles, Norway, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, St Maarten, Sweden, Switzerland, Tajikistan, Turkey, Uganda, United Arab Emirates, United Kingdom, and US Virgin Islands.
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure, Mounds View, MN, United States
- How the firm notified
- Letter
- FDA classified
- Jul 12, 2016
Newest first, one product per recall event; search them all
- MyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The…Z-0399-2018 · Class II · Jul 14, 2017 · Terminated
- Amplia MRI Quad CRT-0 SureScan Claria MRI Quad CRT-D SureScan Compia MRI Quad CRT-D SureScanZ-2117-2017 · Class II · Apr 12, 2017 · Terminated
- Amplia MRI" CRT -D SureScan", Amplia MRI" Quad CRT -D SureScan", Claria MRI" CRT -D SureScan", Claria MRI" Quad…Z-0962-2017 · Class II · Dec 9, 2016 · Terminated
- Medtronic, Evera Implantable Cardioverter Defibrillators: Product Model EVERA XT DDBB1D1, DDBB1D4, DDBB2D1,…Z-2651-2016 · Class II · Aug 12, 2016 · Terminated
- Medtronic, MyCareLink Patient Monitor, Model 24950, Rx Only. The MyCareLink Monitor, Model 24950, is a remote…Z-2125-2016 · Class II · May 26, 2016 · Terminated
- Medtronic CareLink" Monitor (2490C) and Medtronic CareLink Express" Monitor (2020B) Product Usage: The 2490C…Z-1605-2016 · Class II · Mar 31, 2016 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports