Arca
Skip to content
Tools
Tools

Amplia MRI" CRT -D SureScan"

Recall Z-0962-2017, part of FDA recall event 75943. See Medtronic, Cardiac Rhythm and Heart Failure's company page.

Medtronic Inc., Cardiac Rhythm and Heart Failure started this Class II device recall on Dec 9, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Jul 16, 2019.

  • Classification: Class II
  • Started: Dec 9, 2016
  • Status: Terminated

Why It Was Recalled

Class II

During internal testing conducted as part of next generation product development, it was discovered that the AdaptivCRT (aCRT) feature of Amplia MRI and Claria MRI CRT -D devices may appear to be enabled when it is not active.

Record

Product
Amplia MRI" CRT -D SureScan", Amplia MRI" Quad CRT -D SureScan", Claria MRI" CRT -D SureScan", Claria MRI" Quad CRT -D SureScan. implantable medical devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia.
Quantity
21,111 ( 13,198 US; 7,913 OUS).
Lots and codes
Serial Numbers of AMPLIA & CLARIA Released from Manufacturing as of 22-Nov-2016 UPN Serial 00643169543362 RPA200003H 00643169543362 RPA200004H 00643169543362 RPA200005H 00643169543362 RPA200006H 00643169543362 RPA200008H 00643169543362 RPA200009H 00643169543362 RPA200010H 00643169543362 RPA200011H 00643169543362 RPA200012H 00643169543362 RPA200013H 00643169543362 RPA200014H 00643169543362 RPA200015H 00643169543362 RPA200016H 00643169543362 RPA200017H 00643169543362 RPA200018H 00643169543362 RPA200019H 00643169543362 RPA200021H 00643169543362 RPA200022H 00643169543362 RPA200023H 00643169543362…
Distribution
US: All states in USA including DC & PR. OUS: Australia, Finland, Japan, Saudi Arabia, Austria, France, Korea,Singapore, Bangladesh, Germany, Luxembourg, South Africa, Belgium, Greece, Malaysia, Spain, Brunei Darussalam, Guadeloupe, Netherlands, Sweden, Cayman Islands, Hong Kong, New Zealand, Switzerland, Chile, India, Norway, Thailand, Cyprus, Iran, Poland, Turkey, Denmark, Ireland, Portugal, United Arab Emirates, Dominican Republic, Italy, Reunion, United Kingdom.
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure, Mounds View, MN, United States
How the firm notified
Letter
FDA classified
Jan 7, 2017

Medtronic Inc., Cardiac Rhythm and Heart Failure's Other Recalls