Arca
Skip to content
Tools
Tools

Medtronic CareLink" Monitor (2490C) and Medtronic CareLink Express" Monitor (2020B)…

Recall Z-1605-2016, part of FDA recall event 73780. See Medtronic, Cardiac Rhythm and Heart Failure's company page.

Medtronic Inc., Cardiac Rhythm and Heart Failure started this Class II device recall on Mar 31, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Sep 25, 2017.

  • Classification: Class II
  • Started: Mar 31, 2016
  • Status: Terminated

Why It Was Recalled

Class II

A recent firmware update developed by Medtronic for the 2490C CareLink Monitors and 2020B CareLink Express Monitors included incorrect data on the country analog modem dial-up configuration table used to dial into the CareLink Network. This incorrect data resulted in the removal of the 0 prefix necessary for select countries as well as an incorrect alignment of phone numbers for other select cou

Record

Product
Medtronic CareLink" Monitor (2490C) and Medtronic CareLink Express" Monitor (2020B) Product Usage: The 2490C CareLink Home Monitors and 2020B CareLink Express Monitors are remote monitoring systems that interrogate implanted devices and transmit the data to Medtronic s CareLink Network for viewing by the physician. The data is transmitted either through an analog telephone line, a cellular connection or an internet connection.
Quantity
5038
Lots and codes
Model Serial Number 2020B LPH003245P 2020B LPH004235P 2020B LPH004270P 2020B LPH004448P 2020B LPH004578P 2020B LPH006282P 2020B LPH006598P 2020B LPH006892P 2490C IJX010143A 2490C IJX010218A 2490C IJX010291A 2490C IJX010304A 2490C IJX010320A 2490C IJX010414A 2490C IJX010416A 2490C IJX010441A 2490C IJX010474A 2490C IJX010489A 2490C IJX010501A 2490C IJX010521A 2490C IJX014234A 2490C IJX014380A 2490C IJX014382A 2490C IJX014387A 2490C IJX014436A 2490C IJX014473A 2490C IJX014502A 2490C IJX014591A 2490C IJX019468A 2490C IJX019469A 2490C IJX019550A 2490C IJX019568A 2490C IJX019647A 2490C IJX019681A…
Distribution
AUSTRIA, AUSTRALIA, BELGIUM, SWITZERLAND, GERMANY, DENMARK, FRANCE, GREAT BRITAIN, HUNGARY, JAPAN, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, SWEDEN.
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure, Mounds View, MN, United States
How the firm notified
Letter
FDA classified
May 4, 2016

Medtronic Inc., Cardiac Rhythm and Heart Failure's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports