Varian brand Eclipse Treatment Planning System

Recall Z-2462-2012, part of FDA recall event 63093. See Varian Medical Systems Oncology Systems's company page.

Varian Medical Systems, Inc. Oncology Systems started this Class II device recall on Aug 17, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 6, 2014.

  • Classification: Class II
  • Started: Aug 17, 2012
  • Status: Terminated

Why It Was Recalled

Class II

Varian has identified an anomaly with the Eclipse Treatment Planning where an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. Treatment of the patient using these values can therefore lead to under or over-dose.

Record

Product
Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier: CP-08922; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithim is specifically indicated for planning proton treatment of neoplasms of the eyes.
Quantity
6087
Lots and codes
CP-08922: LIST OF SERIAL NUMBERS: H480001, H480130, H480301, H480446, H480623, H480795, H480976, H481088, H481222, H481382, H481544, H481777, H481931, H480002, H480131, H480302, H480448, H480624, H480798, H480981, H481089, H481229, H481383, H481545, H481778, H481932, H480008, H480132, H480305, H480449, H480627, H480808, H480982, H481090, H481230, H481389, H481556, H481787, H481935, H480009, H480133, H480312 ,H480450 ,H480628 ,H480814 ,H480989 ,H481099 ,H481231 ,H481392 ,H481562 ,H481793 ,H481938 ,H480010 ,H480137 ,H480313 ,H480451 ,H480629 ,H480819 ,H480992 ,H481100 ,H481232 ,H481393 ,H481581…
Distribution
Worldwide Distribution - USA (nationwide) and Internationally. . .
Recalling firm
Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA, United States
How the firm notified
Letter
FDA classified
Sep 26, 2012