Varian brand Eclipse Treatment Planning System
Recall Z-2462-2012, part of FDA recall event 63093. See Varian Medical Systems Oncology Systems's company page.
Varian Medical Systems, Inc. Oncology Systems started this Class II device recall on Aug 17, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 6, 2014.
- Classification: Class II
- Started: Aug 17, 2012
- Status: Terminated
Varian has identified an anomaly with the Eclipse Treatment Planning where an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. Treatment of the patient using these values can therefore lead to under or over-dose.
- Product
- Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier: CP-08922; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithim is specifically indicated for planning proton treatment of neoplasms of the eyes.
- Quantity
- 6087
- Lots and codes
- CP-08922: LIST OF SERIAL NUMBERS: H480001, H480130, H480301, H480446, H480623, H480795, H480976, H481088, H481222, H481382, H481544, H481777, H481931, H480002, H480131, H480302, H480448, H480624, H480798, H480981, H481089, H481229, H481383, H481545, H481778, H481932, H480008, H480132, H480305, H480449, H480627, H480808, H480982, H481090, H481230, H481389, H481556, H481787, H481935, H480009, H480133, H480312 ,H480450 ,H480628 ,H480814 ,H480989 ,H481099 ,H481231 ,H481392 ,H481562 ,H481793 ,H481938 ,H480010 ,H480137 ,H480313 ,H480451 ,H480629 ,H480819 ,H480992 ,H481100 ,H481232 ,H481393 ,H481581…
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally. . .
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA, United States
- How the firm notified
- Letter
- FDA classified
- Sep 26, 2012
- Varian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage: The…Z-1163-2013 · Class II · Mar 26, 2013 · Terminated
- Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye…Z-0867-2013 · Class II · Jan 28, 2013 · Terminated
- Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application,…Z-0531-2013 · Class II · Nov 5, 2012 · Terminated
- ARIA Radiation Oncology, Version 11. Varian Medical Systems, Palo Alto, CA. The ARIA Radiation Oncology product is a…Z-0187-2013 · Class II · Sep 24, 2012 · Terminated
- Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA…Z-2383-2012 · Class II · Aug 14, 2012 · Terminated
- Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier:…Z-2153-2012 · Class II · Jul 2, 2012 · Terminated
- Varian brand Varian Eclipse Treatment Planning System for radiotherapy; Model Number: H48; Product is manufactured and…Z-2041-2012 · Class II · Jun 18, 2012 · Terminated
- Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. …Z-1325-2013 · Class II · May 31, 2011 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports