Varian High Energy Clinacs
Recall Z-1325-2013, part of FDA recall event 64997. See Varian Medical Systems Oncology Systems's company page.
Varian Medical Systems, Inc. Oncology Systems started this Class II device recall on May 31, 2011 (voluntary: firm initiated); FDA lists it as terminated, terminated May 15, 2013.
- Classification: Class II
- Started: May 31, 2011
- Status: Terminated
Under certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may result in no more than about 3.5% dose deviation from expected at any point in the beam, and may result in minor injury to the patient or slightly decreased local disease control.
- Product
- Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. Product Usage: The Varian High Energy Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Quantity
- 3943 units
- Lots and codes
- CODES: H140001 H140064 H140137 H140721 H140796 H140862 H260947 H270267 H270428 H270539 H270708 H270828 H271017 H140002 H140065 H140139 H140723 H140797 H140863 H260949 H270271 H270430 H270540 H270709 H270829 H271018 H140003 H140066 H140141 H140724 H140798 H140865 H260950 H270273 H270431 H270542 H270710 H270833 H271019 H140004 H140067 H140143 H140725 H140799 H140866 H260951 H270277 H270435 H270544 H270711 H270836 H271020 H140005 H140068 H140144 H140726 H140800 H140868 H260952 H270280 H270437 H270547 H270712 H270837 H271021 H140006 H140069 H140145 H140727 H140801 H140869 H260953 H270281 H270438…
- Distribution
- Worldwide distribution - USA (nationwide) including Puerto Rico, and countries of: Algeria, Angola, Australia, Austria, Bahrain, Bangladesh, Belarus, Belguim, Brazil, Bosnia and Hezregovina, Bulgaria, Canada, Chile, China, Columbia, Cyprus, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, Frances, French Polynesia, Georgia, Germany, Greece, Gaudaloupe, Gaum, Hong Kong, India, Indonesia, Ireland, Italy, Israel, Latvia, Lithuania, Luxemburg, Japan, Kazarhstan, Kuwait, Lebanon, Macau, Macedonia, Malaysia, Martinique, Mauritania, Mexico, Moldova, Monaco, Morroco, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Phillipines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudia Arabia, Serbia, Singapore, Slovakia, Spain, South Africa, South Korea, Sri Lanka, Syria, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezula, Vietnam, and Zimbabwe.
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA, United States
- How the firm notified
- Letter
- FDA classified
- May 14, 2013
- Varian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage: The…Z-1163-2013 · Class II · Mar 26, 2013 · Terminated
- Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye…Z-0867-2013 · Class II · Jan 28, 2013 · Terminated
- Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application,…Z-0531-2013 · Class II · Nov 5, 2012 · Terminated
- ARIA Radiation Oncology, Version 11. Varian Medical Systems, Palo Alto, CA. The ARIA Radiation Oncology product is a…Z-0187-2013 · Class II · Sep 24, 2012 · Terminated
- Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier:…Z-2462-2012 · Class II · Aug 17, 2012 · Terminated
- Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA…Z-2383-2012 · Class II · Aug 14, 2012 · Terminated
- Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier:…Z-2153-2012 · Class II · Jul 2, 2012 · Terminated
- Varian brand Varian Eclipse Treatment Planning System for radiotherapy; Model Number: H48; Product is manufactured and…Z-2041-2012 · Class II · Jun 18, 2012 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports