Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition
Recall Z-0867-2013, part of FDA recall event 64319. See Varian Medical Systems Oncology Systems's company page.
Varian Medical Systems, Inc. Oncology Systems started this Class II device recall on Jan 28, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 4, 2014.
- Classification: Class II
- Started: Jan 28, 2013
- Status: Terminated
The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.
- Product
- Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
- Quantity
- 1168 units
- Lots and codes
- H480044 , H481353 , H482733 , H484585 , H485948 , H480045 , H481354 , H482734 , H484634 , H486024 , H480105 , H481433 , H482765 , H484673 , H486032 , H480106 , H481436 , H482827 , H484677 , H486093 , H480159 , H481437 , H482858 , H484680 , H486094 , H480160 , H481451 , H482859 , H484745 , H486095 , H480247 , H481452 , H482860 , H484746 , H486114 , H480301 , H481453 , H482888 , H484747 , H486140 , H480373 , H481454 , H482889 , H484753 , H486209 , H480374 , H481460 , H482897 , H484830 , H486291 , H480378 , H481461 , H483087 , H484842 , H486369 , H480379 , H481462 , H483204 , H484856 , H486385 ,…
- Distribution
- Worldwide Distribution - USA Nationwide
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA, United States
- How the firm notified
- Letter
- FDA classified
- Feb 20, 2013
Newest first, one product per recall event; search them all
- Varian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage: The…Z-1163-2013 · Class II · Mar 26, 2013 · Terminated
- Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application,…Z-0531-2013 · Class II · Nov 5, 2012 · Terminated
- ARIA Radiation Oncology, Version 11. Varian Medical Systems, Palo Alto, CA. The ARIA Radiation Oncology product is a…Z-0187-2013 · Class II · Sep 24, 2012 · Terminated
- Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier:…Z-2462-2012 · Class II · Aug 17, 2012 · Terminated
- Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA…Z-2383-2012 · Class II · Aug 14, 2012 · Terminated
- Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier:…Z-2153-2012 · Class II · Jul 2, 2012 · Terminated
- Varian brand Varian Eclipse Treatment Planning System for radiotherapy; Model Number: H48; Product is manufactured and…Z-2041-2012 · Class II · Jun 18, 2012 · Terminated
- Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. …Z-1325-2013 · Class II · May 31, 2011 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports