Varian brand Clinac, Trilogy
Recall Z-2383-2012, part of FDA recall event 63033. See Varian Medical Systems Oncology Systems's company page.
Varian Medical Systems, Inc. Oncology Systems started this Class II device recall on Aug 14, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 4, 2013.
- Classification: Class II
- Started: Aug 14, 2012
- Status: Terminated
Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.
- Product
- Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. 1) The Varian Low Energy Linear Accelerator is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment i is indicated. (2) The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Quantity
- 221 Units
- Lots and codes
- LIST OF SERIAL NUMBERS: H140863 ,H294608 ,H294669 ,H294732 ,H140865 ,H294610 ,H294670 ,H294733 ,H140866 ,H294611 ,H294671 ,H294734 ,H140867 ,H294612 ,H294672 ,H294735 ,H140868 ,H294613 ,H294673 ,H294736 ,H140869 ,H294614 ,H294674 ,H294737 ,H140870 ,H294615 ,H294675 ,H294738 ,H140871 ,H294616 ,H294676 ,H294739 ,H140872 ,H294617 ,H294677 ,H294740 ,H140873 ,H294618 ,H294678 ,H294741 ,H140874 ,H294619 ,H294679 ,H294742 ,H140875 ,H294620 ,H294680 ,H294743 ,H140876 ,H294621 ,H294681 ,H294744 ,H140877 ,H294622 ,H294682 ,H294745 ,H140878 ,H294623 ,H294683 ,H294746 ,H140879 ,H294624 ,H294684 ,H294747…
- Distribution
- Worldwide Distribution.
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA, United States
- How the firm notified
- Letter
- FDA classified
- Sep 14, 2012
- Varian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage: The…Z-1163-2013 · Class II · Mar 26, 2013 · Terminated
- Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye…Z-0867-2013 · Class II · Jan 28, 2013 · Terminated
- Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application,…Z-0531-2013 · Class II · Nov 5, 2012 · Terminated
- ARIA Radiation Oncology, Version 11. Varian Medical Systems, Palo Alto, CA. The ARIA Radiation Oncology product is a…Z-0187-2013 · Class II · Sep 24, 2012 · Terminated
- Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier:…Z-2462-2012 · Class II · Aug 17, 2012 · Terminated
- Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier:…Z-2153-2012 · Class II · Jul 2, 2012 · Terminated
- Varian brand Varian Eclipse Treatment Planning System for radiotherapy; Model Number: H48; Product is manufactured and…Z-2041-2012 · Class II · Jun 18, 2012 · Terminated
- Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. …Z-1325-2013 · Class II · May 31, 2011 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports