Varian brand Varian Eclipse Treatment Planning System for radiotherapy; Model Number:…

Recall Z-2041-2012, part of FDA recall event 62420. See Varian Medical Systems Oncology Systems's company page.

Varian Medical Systems, Inc. Oncology Systems started this Class II device recall on Jun 18, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 6, 2014.

  • Classification: Class II
  • Started: Jun 18, 2012
  • Status: Terminated

Why It Was Recalled

Class II

Varian has identified an anomaly with the Eclipse Treatment Planning (for radiotherapy treatments) where changing the prescription in a calculated VMAT or RapidArc plan may lead to Monitor Units which do not reflect the new prescription. If the dose prescription of a VMAT or RapidArc plan is changed after the initial Dose Volume Optimization, the Monitor Units and dose distribution may not reflect

Record

Product
Varian brand Varian Eclipse Treatment Planning System for radiotherapy; Model Number: H48; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Quantity
7113 units
Lots and codes
Affected Devices: H480110 ,H486186 ,H48B178 ,H48D026 ,H48D966 ,H487806 ,H48B764 ,H48B218 ,H48C503 ,H488467 ,H481815 ,H48D923 ,H48E486 ,H480267 ,H486446 ,H48B179 ,H48D027 ,H48A565 ,H487807 ,H48B765 ,H48B963 ,H48B756 ,H488468 ,H481816 ,H48E896 ,H48B426 ,H480268 ,H486447 ,H48B180 ,H48D028 ,H48A566 ,H487808 ,H482178 ,H48B964 ,H48B757 ,H489917 ,H485467 ,H48E897 ,H48B668 ,H480269 ,H487164 ,H48B181 ,H48D029 ,H48D566 ,H487809 ,H489962 ,H48B965 ,H482450 ,H48A912 ,H48A254 ,H489213 ,H48B669 ,H480270 ,H487165 ,H48B182 ,H48D030 ,H488558 ,H487810 ,H489963 ,H48B966 ,H482451 ,H48A913 ,H48C429 ,H489214…
Distribution
Worldwide distribution: USA (nationwide) and countries including of: Austria, Australia, Bahrain, Belgium, Brazil, Canada, China, Chile, Colombia, Denmark, Egypt, Finland, France,Germany, Greece, Hungary, Hong Kong, India, Indonesia, Israel, Ireland, Italy, Japan, Latvia, Malaysia, Mexico, Morocco, Nepal, Netherlands, Norway, Panama, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, United Kingdom, Taiwan, Turkey, Ukraine, United Arab Emirates and Venezuela.
Recalling firm
Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA, United States
How the firm notified
Letter
FDA classified
Jul 19, 2012