Recall Z-0891-2022, part of FDA recall event 89794. See Abbott Laboratories's company page.
Abbott Laboratories started this Class II device recall on Mar 8, 2022 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Mar 8, 2022
- Status: Ongoing
This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).
- Product
- ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.
- Quantity
- 402 kits
- Lots and codes
- UDI: (01)00380740003272(17)220531(10)166000
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) including countries of: AUSTRIA, BELGIUM, BOSNIA AND HERZEGOVI, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, GREECE, IRAQ, ITALY, KAZAKHSTAN, LITHUANIA, LUXEMBOURG, MACEDONIA, MOROCCO, PANAMA, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, and TURKEY
- Recalling firm
- Abbott Laboratories, Abbott Park, IL, United States
- How the firm notified
- —
- FDA classified
- Apr 7, 2022
Newest first, one product per recall event; search them all
- Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20Z-0961-2026 · Class II · Nov 13, 2025 · Ongoing
- Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version:…Z-0489-2026 · Class II · Sep 18, 2025 · Ongoing
- Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-03 Software Version: All software…Z-1653-2025 · Class II · Mar 18, 2025 · Ongoing
- Alinity hq Analyzer, REF: 09P68-01Z-2681-2024 · Class II · Jun 14, 2024 · Ongoing
- ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.Z-1476-2024 · Class II · Feb 21, 2024 · Ongoing
- Alinity c Processing Module REF 03R67-01 The Alinity c Processing Module is a fully automated chemistry analyzer…Z-0681-2024 · Class II · Oct 19, 2023 · Ongoing
- Abbott CELL-DYN Ruby, Model CD-Ruby, List number 08H67-01, and Abbott CELL-DYN Sapphire, Model CD Sapphire, list number…Z-0604-2024 · Class II · Oct 19, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports