BATTERY FLEX-3S2P 10.8V 18650 LI-ION SMBUS used with affected patient monitors Model No…
Recall Z-0457-2026, part of FDA recall event 97786. See GE Healthcare Finland's company page.
GE Healthcare Finland Oy started this Class II device recall on Sep 26, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Sep 26, 2025
- Status: Ongoing
- Products in this event: 5
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
- Product
- BATTERY FLEX-3S2P 10.8V 18650 LI-ION SMBUS used with affected patient monitors Model No 2062895-001 Manufacturer Part Number U80296-1R01, revision S
- Quantity
- 2857
- Lots and codes
- No UDI Manufacturing date range (YYYY-MM-DD date format, located underneath date of manufacture symbol): 2024-11-25 - 2025-05-31; 2025-07-07 Serial Nos. 021725-20339 021725-22682 120924-06125 120924-05011 120924-05830 120924-03583 120924-04467 120924-04447 031425-28687 120924-04436 120924-04240 021725-21549 021725-20604 5567418-SO-1-1 031425-30157 031425-28597 031425-29489 031425-28901 031425-28978 031425-28590 031425-29484 031425-29177 021725-21521 021725-20357 021725-20412 021725-22331 120924-04465 120924-07616 120924-07484 120924-07792 120924-07909 120924-07359 120924-07524 120924-06997…
- Distribution
- Worldwide - US Nationwide distribution.
- Recalling firm
- GE Healthcare Finland Oy, Helsinki, Finland
- How the firm notified
- Letter
- FDA classified
- Nov 7, 2025
- B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005…Z-0455-2026 · Class II · Sep 26, 2025 · Ongoing
- CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a…Z-0454-2026 · Class II · Sep 26, 2025 · Ongoing
- Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or AlarmsZ-0456-2026 · Class II · Sep 26, 2025 · Ongoing
- Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a…Z-0453-2026 · Class II · Sep 26, 2025 · Ongoing