CARESCAPE Canvas Smart Display
Recall Z-2169-2024, part of FDA recall event 94494. See GE Healthcare Finland's company page.
GE Healthcare Finland Oy started this Class II device recall on Apr 5, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Apr 5, 2024
- Status: Ongoing
- Products in this event: 5
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
- Product
- CARESCAPE Canvas Smart Display, Model Numbers: a) 5700023, b) 5700023-02, c) 5939590-102; monitor, physiological, patient
- Quantity
- 4649 units
- Lots and codes
- GTIN 00195278522313: a) 5700023, Serial Numbers: 2534526-3-1, 2553213-5-1, 2606854-2-1, 5220256-1-1, 5215673-1-1, 5215634-1-1, 5264236-1-1, 5264113-1-1, 5245953-1-1, 5264089-1-1, 5275734-1-1, 5301535-1-1, 5300969-1-1, 5341313-1-1, 5395275-1-1, 5232613-1-1, 5232595-1-1, 5287222-1-1, 5390098-3-1, 5316614-6-1, 5316614-16-1, 5316614-22-1, 2615004-27-1, 5368586-1-1, 5223040-1-1, 5223040-1-2, 5223040-1-3, 5223040-1-4, 5274472-12-1, 5274472-71-1, 5350377-1-1, 5215675-1-1, 5223119-1-1, 5231395-1-1, 5245547-1-1, 5307272-2-1, 5318051-3-1, 5318978-2-1, 5317879-3-1, 5331606-1-1, 5338491-2-1, 5356848-4-1,…
- Distribution
- Worldwide distribution.
- Recalling firm
- GE Healthcare Finland Oy, Helsinki, Finland
- How the firm notified
- Letter
- FDA classified
- Jun 21, 2024
- CARESCAPE Canvas 1000, Model Numbers: a) 5865770; b) 5865770-01040169; c) 5865770-01054142; d) 5865770-01054696; …Z-2168-2024 · Class II · Apr 5, 2024 · Ongoing
- CARESCAPE Canvas Smart Display, Model Numbers: a) 5514418, b) 5514418 / MAEP2, c) 5514418-01059642, d)…Z-2170-2024 · Class II · Apr 5, 2024 · Ongoing
- CARESCAPE B850 Model Numbers: a) MBC303, Model Numbers: 1) 2078633-001, 2) 2078633-010, 3) 2078633-012, 4)…Z-2172-2024 · Class II · Apr 5, 2024 · Ongoing
- CARESCAPE B650: Model Numbers: a) MBB313, Model Numbers: 1) 2095801-001, 2) 2095801-001-01018416, 3)…Z-2171-2024 · Class II · Apr 5, 2024 · Ongoing
- Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a…Z-0453-2026 · Class II · Sep 26, 2025 · Ongoing
- TruSignal AllFit Sensor: a) REF TS-AF-10, 10 pcs; B) REF AF-25, 25 pcs; OximeterZ-2027-2023 · Class I · May 19, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports