Arca
Skip to content
Tools
Tools

CARESCAPE B850 Model Numbers: a) MBC303

Recall Z-2172-2024, part of FDA recall event 94494. See GE Healthcare Finland's company page.

GE Healthcare Finland Oy started this Class II device recall on Apr 5, 2024 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Apr 5, 2024
  • Status: Ongoing
  • Products in this event: 5

Why It Was Recalled

Class II

GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.

Record

Product
CARESCAPE B850 Model Numbers: a) MBC303, Model Numbers: 1) 2078633-001, 2) 2078633-010, 3) 2078633-012, 4) 2078633-017, 5) 2078633-028, 6) 2095802-001-54069330, 7) 2095802-001-554922, 8) 2095802-001-554923, 9) 2095802-001-554924, 10) 2095802-001-554925, 11) 2095802-001-554926, 12) 2095802-001-554927, 13) 2095802-001-554928, 14) 2095802-001-555120, 15) 2095802-001-556223, 16) 2095802-001-557346, 17) 2095802-001-558678, 18) 2095802-001-558707, 19) 2095802-001-559208, 20) 2095802-001-559337, 21) 2095802-001-559916, 22) 2095802-001-560446, 23) 2095802-001-560519, 24) 2095802-001-560520, 25) 2095802-001-560521, 26) 2095802-001-560522, 27) 2095802-001-560637, 28) 2095802-001-560803, 29) 2095802-001-561557, 30) 2095802-001-562090, 31) 2095802-001-562136, 32) 2095802-001-562249, 33) 2095802-001-562799, 34) 2095802-001-563317, 35) 2095802-001-563770, 36) 2095802-001-563771, 37) 2095802-001-564033, 38) 2095802-001-564174, 39) 2095802-001-565603, 40) 2095802-001-566307, 41) 2095802-001-566308, 42) 2095802-001-566696, 43) 2095802-001-566798, 44) 2095802-001-566816, 45) 2095802-001-566825, 46) 2095802-001-566826, 47) 2095802-001-566925, 48) 2095802-001-567844, 49) 2095802-001-568125, 50) 2095802-001-568726, 51) 2095802-001-568765, 52) 2095802-001-569226, 53) 2095802-001-569235, 54) 2095802-001-569318, 55) 2095802-001-569322, 56) 2095802-001-569329, 57) 2095802-001-569332, 58) 2095802-001-569426, 59) 2095802-001-569860, 60) 2095802-001-569871, 61) 2095802-001-569990, 62) 2095802-001-570062, 63) 2095802-001-570185, 64) 2095802-001-570746, 65) 2095802-001-570960, 66) 2095802-001-571048, 67) 2095802-001-571049, 68) 2095802-001-571801, 69) 2095802-001-571807, 70) 2095802-001-572988, 71) 2095802-001-572991, 72) 2095802-001-573205, 73) 2095802-001-574167, 74) 2095802-001-574403, 75) 2095802-001-574569, 76) 2095802-001-574570, 77) 2095802-001-574685, 78) 2095802-001-575604, 79) 2095802-001-575842, 80) 2095802-001-576569, 81) 2095802-001-577254, 82) 2095802-001-577260, 83) 2095802-001-577472, 84) 2095802-001-577688, 85) 2095802-001-578467, 86) 2095802-001-578468, 87) 2095802-001-579029, 88) 2095802-001-579134, 89) 2095802-001-579366, 90) 2095802-001-579526, 91) 2095802-001-579590, 92) 2095802-001-579604, 93) 2095802-001-579692, 94) 2095802-001-579926, 95) 2095802-001-579933, 96) 2095802-001-580487, 97) 2095802-001-580731, 98) 2095802-001-581010, 99) 2095802-001-581094, 100) 2095802-001-581417, 101) 2095802-001-581433, 102) 2095802-001-582510, 103) 2095802-001-582710, 104) 2095802-001-582906, 105) 2095802-001-582907, 106) 2095802-001-583150, 107) 2095802-001-583551, 108) 2095802-001-583553, 109) 2095802-001-583888, 110) 2095802-001-583970, 111) 2095802-001-584288, 112) 2095802-001-584547, 113) 2095802-001-584810, 114) 2095802-001-585047, 115) 2095802-001-585468, 116) 2095802-001-585907, 117) 2095802-001-586028, 118) 2095802-001-586829, 119) 2095802-001-587286, 120) 2095802-001-587726, 121) 2095802-001-587767, 122) 2095802-001-587770, 123) 2095802-001-588006, 124) 2095802-001-588111, 125) 2095802-001-588571, 126) 2095802-001-589110, 127) 2095802-001-589168, 128) 2095802-001-589304, 129) 2095802-001-589326, 130) 2095802-001-589329, 131) 2095802-001-589511, 132) 2095802-001-589655, 133) 2095802-001-589796, 134) 2095802-001-589799, 135) 2095802-001-589801, 136) 2095802-001-589993, 137) 2095802-001-590263, 138) 2095802-001-590270, 139) 2095802-001-590640, 140) 2095802-001-590643, 141) 2095802-001-590645, 142) 2095802-001-590648, 143) 2095802-001-590649, 144) 2095802-001-590651, 145) 2095802-001-590652, 146) 2095802-001-591048, 147) 2095802-001-591102, 148) 2095802-001-591244, 149) 2095802-001-591246, 150) 2095802-001-591718, 151) 2095802-001-592666, 152) 2095802-001-592669, 153) 2095802-001-592683, 154) 2095802-001-592983, 155) 2095802-001-593000, 156) 2095802-001-593024, 157) 2095802-001-593039,
Quantity
25268 units
Lots and codes
a) model MBC303: 1) 2078633-001, UDI/DI none, Serial Numbers: 2293663-1-1; 2) 2078633-010, UDI/DI none, Serial Numbers: 2284163-2-1, 2284112-3-1, 2284163-1-1, 2284112-4-1, 2262484-1-1, 2262484-1-2; 3) 2078633-012, UDI/DI none, Serial Numbers: 2400553-4-1, 2266860-3-2; 4) 2078633-017, UDI/DI none, Serial Numbers: 2261264-2-1, 2290421-2-1, 2288663-1-1, 2290422-2-1, 2312000-3-1, 2311672-4-1, 2311672-5-1, 2320320-1-1, 2284038-3-1, 2329280-2-1; 5) 2078633-028, UDI/DI none, Serial Numbers: 2426418-2-1, 2450634-2-1, 2501447-2-1, 2426426-25-1, 2428347-1-1, 2460598-2-1, 2405070-10-1, 2426421-3-1,…
Distribution
Worldwide distribution.
Recalling firm
GE Healthcare Finland Oy, Helsinki, Finland
How the firm notified
Letter
FDA classified
Jun 21, 2024

Other Products in This Recall

Recall event 94494

GE Healthcare Finland Oy's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports