Arca
Skip to content
Tools
Tools

Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or…

Recall Z-0453-2026, part of FDA recall event 97786. See GE Healthcare Finland's company page.

GE Healthcare Finland Oy started this Class II device recall on Sep 26, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Sep 26, 2025
  • Status: Ongoing
  • Products in this event: 5

Why It Was Recalled

Class II

Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.

Record

Product
Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/26 to include: CARESCAPE B450 MBA313 2095800-001 CARESCAPE MONITOR B450 (ATO MODEL) 2068491-002 CARESCAPE MONITOR B450 B1 2107633-001 CARESCAPE MONITOR B450-GS (V2.0.8) 2094080-001
Quantity
3271
Lots and codes
UDI-DI 00840682146135 Serial numbers SNT25079613SA SNT25079608SA SNT25079622SA SNT25079618SA SNT25079619SA SNT25079614SA SNT25079611SA SNT25079621SA SNT25079620SA SNT25079603SA SNT25079616SA SNT25079612SA SNT25079602SA SNT25079615SA SNT25079663SA SNT25079664SA SNT25079677SA SNT25079662SA SNT25079667SA SNT25079678SA SNT25079680SA SNT25079666SA SNT25079695SA SNT25079694SA SNT25079697SA SNT25079688SA SNT25079687SA SNT25079685SA SNT25079676SA SNT25079674SA SNT25079671SA SNT25079691SA SNT25079689SA SNT25079679SA SNT25079681SA SNT25079684SA SNT25079672SA SNT25079668SA SNT25079665SA SNT25079675SA…
Distribution
Worldwide - US Nationwide distribution.
Recalling firm
GE Healthcare Finland Oy, Helsinki, Finland
How the firm notified
Letter
FDA classified
Nov 7, 2025

Other Products in This Recall

Recall event 97786

GE Healthcare Finland Oy's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc