CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or…
Recall Z-0454-2026, part of FDA recall event 97786. See GE Healthcare Finland's company page.
GE Healthcare Finland Oy started this Class II device recall on Sep 26, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Sep 26, 2025
- Status: Ongoing
- Products in this event: 5
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
- Product
- CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Updated 3/19 to add CARESCAPE ONE MBZ101 2087075-001
- Quantity
- 5935
- Lots and codes
- UDI-DI 00195278288639 Serial numbers SV925270018SA SV925270014SA SV925270015SA SV925260105SA SV925270020SA SV925270013SA SV925270017SA SV925270016SA SV925060394SA SV925080329SA SV925080345SA SV925080323SA SV925080348SA SV925080153SA SV925080340SA SV925080281SA SV925080346SA SV925290119SA SV925290090SA SV925180134SA SV925160111SA SV925180131SA SV925180139SA SV925180140SA SV925180142SA SV925180143SA SV925180151SA SV925180152SA SV925160114SA SV925180141SA SV925180126SA SV925180132SA SV925180136SA SV925180135SA SV925180144SA SV925180146SA SV925180149SA SV925270007SA SV925260094SA SV925270009SA…
- Distribution
- Worldwide - US Nationwide distribution.
- Recalling firm
- GE Healthcare Finland Oy, Helsinki, Finland
- How the firm notified
- Letter
- FDA classified
- Nov 7, 2025
- B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005…Z-0455-2026 · Class II · Sep 26, 2025 · Ongoing
- Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or AlarmsZ-0456-2026 · Class II · Sep 26, 2025 · Ongoing
- BATTERY FLEX-3S2P 10.8V 18650 LI-ION SMBUS used with affected patient monitors Model No 2062895-001 Manufacturer…Z-0457-2026 · Class II · Sep 26, 2025 · Ongoing
- Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a…Z-0453-2026 · Class II · Sep 26, 2025 · Ongoing
Newest first, one product per recall event; search them all
- CARESCAPE Canvas Smart Display, Model Numbers: a) 5700023, b) 5700023-02, c) 5939590-102; monitor, physiological,…Z-2169-2024 · Class II · Apr 5, 2024 · Ongoing
- TruSignal AllFit Sensor: a) REF TS-AF-10, 10 pcs; B) REF AF-25, 25 pcs; OximeterZ-2027-2023 · Class I · May 19, 2023 · Ongoing