B105P, B125P, B105M, B125M
Recall Z-0455-2026, part of FDA recall event 97786. See GE Healthcare Finland's company page.
GE Healthcare Finland Oy started this Class II device recall on Sep 26, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Sep 26, 2025
- Status: Ongoing
- Products in this event: 5
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
- Product
- B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005 6160000-101 6160000-102 6160000-103 6160000-104 6160000-105 6160001-063 6160001-064 6160001-065 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms 3/19/26 updated to add B105P Patient Monitor 6160001-061 B125P Patient Monitor 6160001-062 B105M-C Patient Monitor 6160002-033 B105M-A Patient Monitor 6160000-113 B155M-A Patient Monitor 6160000-115 B125P-B Patient Monitor 6160002-022 B105M-B Patient Monitor 6160002-023 B125M-B Patient Monitor 6160002-024 B105M-OR Patient Monitor 6160000-143 B105M-OR Patient Monitor 6160000-144 B155M-OR Patient Monitor 6160000-145
- Quantity
- 7408
- Lots and codes
- Model No. 6160000-002 UDI-DI 00840682147224 Serial Nos. SR525270031WA SR525270032WA SR525270033WA SR525270034WA SR525270035WA SR525270036WA SR525270037WA SR525270038WA SR525270039WA SR525270040WA SR525270041WA SR525270042WA SR525270043WA SR525270044WA SR525270045WA SR525270046WA SR525270047WA SR525270048WA SR525270049WA SR525270050WA SR525270051WA SR525270052WA SR525270053WA SR525270054WA SR525270055WA SR525270056WA SR525270057WA SR525270058WA SR525270059WA SR525270060WA ; Model No. 6160000-003 UDI-DI 00840682146708 Serial Nos. SR625210256SA SR625200211SA SR625200212SA SR625200213SA…
- Distribution
- Worldwide - US Nationwide distribution.
- Recalling firm
- GE Healthcare Finland Oy, Helsinki, Finland
- How the firm notified
- Letter
- FDA classified
- Nov 7, 2025
- CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a…Z-0454-2026 · Class II · Sep 26, 2025 · Ongoing
- Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or AlarmsZ-0456-2026 · Class II · Sep 26, 2025 · Ongoing
- BATTERY FLEX-3S2P 10.8V 18650 LI-ION SMBUS used with affected patient monitors Model No 2062895-001 Manufacturer…Z-0457-2026 · Class II · Sep 26, 2025 · Ongoing
- Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a…Z-0453-2026 · Class II · Sep 26, 2025 · Ongoing
Newest first, one product per recall event; search them all
- CARESCAPE Canvas Smart Display, Model Numbers: a) 5700023, b) 5700023-02, c) 5939590-102; monitor, physiological,…Z-2169-2024 · Class II · Apr 5, 2024 · Ongoing
- TruSignal AllFit Sensor: a) REF TS-AF-10, 10 pcs; B) REF AF-25, 25 pcs; OximeterZ-2027-2023 · Class I · May 19, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports