ARTIS pheno

Interventional Fluoroscopic X-Ray System. 510(k) K163286, from Siemens Medical Solution USA, Inc. (Malvern, PA). Product code OWB, class 2.

As of Nov 18, 2026, K163286 is an FDA 510(k) held by Siemens Medical Solution USA, Inc., substantially equivalent on Mar 9, 2017 after 107 days, with 14 recalls tied to it in FDA's recall database.

  • Review time: 107 days
  • Recalls: 14
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate
  • ARTIS pheno (VE2)K201156 · Siemens Medical Solution USA, Inc. · Jun 29, 2020
  • ARTIS iconoK190768 · Siemens Medical Solutions USA, Inc. · Sep 12, 2019
  • MOBILETT Elara MaxK182639 · Siemens Medi Cal Solutions, Inc. · Mar 15, 2019
  • Cios AlphaK181560 · Siemens Medical Systems USA, Inc. · Oct 24, 2018
  • Cios SpinK181550 · Siemens Medical Solutions USA, Inc. · Oct 30, 2018

Recalls

Recalls FDA ties to K163286, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 2, 2023Z-1498-2023If, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range the…Open
Feb 2, 2023Z-1311-2023In the event of any unintended table movement, the system may not detect the incorrect direction, could lead to the injury of a patient, staff member, operator, or equipment.Open
Dec 16, 2022Z-0956-2023Mechanical connection between the tabletop and table base may be lost.Open
Sep 20, 2022Z-0119-2023Siemens has become aware of three potential software issues with ARTIS pheno systems with software version VE10B. This may lead to a hazardous situation for patients if treatment cannot be continued on the system and…Open
Feb 25, 2022Z-1339-2022After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient orientation/position may be misinterpreted and result in…Open
May 27, 2021Z-2029-2021Software error VE10B -In case of a fault within the drivetrain of an axis and while performing a stepped peripheral imaging (PERISTEPPING/PERIVISION), the C-arm may leave its intended travel path by up to 22 cm (9")…Open
Apr 22, 2020Z-2108-2020If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible,…Terminated
Jan 24, 2020Z-1340-2020Potential exists for one or more axes of the robotic stand to become blocked by their safety brakes; In the event the axis is blocked by its safety brake, the robotic stand can no longer perform the intended movementsTerminated
Dec 20, 2019Z-0807-2020The potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred.Terminated
Jun 6, 2019Z-1905-2019Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted…Terminated
Jul 29, 2018Z-3190-2018Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for…Terminated
Jun 22, 2018Z-2748-2018Potential hardware issue which may cause thermal effects and possible damage to system parts.Terminated
May 30, 2018Z-2524-2018During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a…Terminated
May 1, 2018Z-2283-2018If the C-arm is manually moved in the longitudinal direction to a fully extended position, the system's safety stop mechanism may be activated. As a result, the C-arm stops the movement and can drop up to 20 mm (0.8 ).…Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Nov 22, 2016
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.