Tools

Cios Spin

Interventional Fluoroscopic X-Ray System. 510(k) K181550, from Siemens Medical Solutions USA, Inc. (Malvern, PA). Product code OWB, class 2.

As of Nov 18, 2026, K181550 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Oct 30, 2018 after 139 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 139 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Primary Predicate: Cios Alpha -System for Image Acquisition Product Codes: OWB, OXO -Post processing Software -Detector Siemens K163286 March 09, 2017 -Anti-Microbial Coating Secondary Predicate: ARTIS -Ant-Microbial Coating Claims pheno Produ ct Codes: OWB, JAA -Cleaning Instructions Siemens K051133 June 1, 2005 -Digital Cine Mode (DCM) Primary Predicate: ARCADIS Avantic Product Codes: OXO…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K181550, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 30, 2020Z-2888-2020Siemens has become aware of a hardware error which affects Cios Spin and Cios Alpha systems after the previous repair of the X-ray Generator with an integrated Energy Storage Unit (ESU).Terminated
Jul 14, 2020Z-2848-2020Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of…Terminated
Apr 20, 2020Z-1959-2020Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the…Terminated
Apr 17, 2020Z-2128-2020The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other…Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Jun 13, 2018
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files