Arca
Skip to content
Tools
Tools

Cios Alpha

Interventional Fluoroscopic X-Ray System. 510(k) K132094, from Siemens Medi Cal Solutions, Inc. (Mavern, PA). Product code OWB, class 2.

As of Nov 21, 2026, K132094 is an FDA 510(k) held by Siemens Medi Cal Solutions, Inc., substantially equivalent on Mar 11, 2014 after 246 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 246 days
  • Recalls: 8
  • Cited as a predicate by: 5 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • Arcadis AvanticK051133 · Siemens Medical Solutions USA, Inc. · Jun 1, 2005

Cited as a predicate

5 later 510(k) summaries name this device as a predicate
  • Cios Alpha; Cios FlowK251520 · Siemens Medical Solutions USA, Inc. · Jul 9, 2025
  • Cios AlphaK181560 · Siemens Medical Systems USA, Inc. · Oct 24, 2018
  • Cios SpinK181550 · Siemens Medical Solutions USA, Inc. · Oct 30, 2018
  • Cios FusionK153244 · Siemens Medical Solutions USA, Inc. · Mar 7, 2016
  • Cios ConnectK153147 · Siemens Medical Solutions USA, Inc. · Jan 12, 2016

Recalls

Recalls FDA ties to K132094, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 14, 2020Z-2848-2020Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of…Terminated
Nov 25, 2019Z-0782-2020During a routine inspection of a system, a crack on the C-arm holder was discovered. In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformations of the C-arm.…Terminated
Jun 20, 2019Z-1953-2019DAP chamber may be missing an insulating foilTerminated
Mar 18, 2019Z-1298-2019Continuous operation at very high tube output may result in increased wear of the radiation emitting components. This may occur when tube voltages of 125 kV are used for a long period, e.g. 30 minutes or more of…Terminated
Feb 19, 2016Z-1278-2016Software issues on Cios Alpha mobile C-Arm systemTerminated
Feb 16, 2016Z-1281-2016Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is…Terminated
Aug 17, 2015Z-0118-2016patient procedure interruption due to a potential system failureTerminated
May 5, 2015Z-1958-2015under certain circumstances the Cios Alpha system may freeze during a procedure.Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Jul 8, 2013
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.