Cios Alpha
Interventional Fluoroscopic X-Ray System. 510(k) K181560, from Siemens Medical Systems USA, Inc. (Malvern, PA). Product code OWB, class 2.
As of Nov 18, 2026, K181560 is an FDA 510(k) held by Siemens Medical Systems USA, Inc., substantially equivalent on Oct 24, 2018 after 132 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 132 days
- Recalls: 7
- Cited as a predicate by: 2 devices
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Cios AlphaK132094 · Siemens Medi Cal Solutions, Inc. · Mar 11, 2014
- ARTIS phenoK163286 · Siemens Medical Solution USA, Inc. · Mar 9, 2017
- Arcadis AvanticK051133 · Siemens Medical Solutions USA, Inc. · Jun 1, 2005
Cited as a predicate
2 later 510(k) summaries name this device as a predicate- Cios AlphaK210055 · Siemens Medical Solutions USA, Inc. · Feb 5, 2021
- Cios FlowK203504 · Siemens Medical Solutions USA, Inc. · Dec 22, 2020
Recalls
Recalls FDA ties to K181560, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jul 30, 2020 | Z-2887-2020Siemens has become aware of a hardware error which affects Cios Spin and Cios Alpha systems after the previous repair of the X-ray Generator with an integrated Energy Storage Unit (ESU). | Terminated |
| Jul 14, 2020 | Z-2848-2020Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of… | Terminated |
| Apr 20, 2020 | Z-1958-2020, Z-1957-2020Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the… | Terminated |
| Apr 17, 2020 | Z-2127-2020The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other… | Terminated |
| Jun 20, 2019 | Z-1953-2019DAP chamber may be missing an insulating foil | Terminated |
| Mar 22, 2019 | Z-1295-2019Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists | Terminated |
Other OWB clearances
The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device- uAngio AVIVA CXK253726 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 21, 2026
- Extron 3;EXTRON 6K261413 · DRTECH Corporation · Aug 19, 2026
- Zenition 30K254205 · Philips Medical Systems Nederland B.V. · Aug 7, 2026
- Shield System (Siemens 30x40)K260607 · Radiaction , Ltd. · Jul 31, 2026
- SmartCT (R3.0 (L10))K253670 · Philips Medical Systems B.V. · Jul 10, 2026
- C-beamK252735 · Pulmera, Inc. · May 22, 2026
- Intelligent NRK252503 · Canon, Inc. · Apr 30, 2026
- ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoK254173 · Siemens Medical Solutions USA, Inc. · Apr 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 14, 2018
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files