Oec 3D
Image-Intensified Fluoroscopic X-Ray System, Mobile. 510(k) K203346, from Ge Oec Medical Systems, Inc. (Salt Lake City, UT). Product code OXO, class 2.
As of Nov 17, 2026, K203346 is an FDA 510(k) held by Ge Oec Medical Systems, Inc., substantially equivalent on Mar 5, 2021 after 112 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 5, 2021
- Review time: 112 days
- Recalls: 3
- Cited as a predicate by: 2 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- OEC EliteK192819 · Ge Oec Medical Systems, Inc. · Nov 8, 2019
- Cios SpinK181550 · Siemens Medical Solutions USA, Inc. · Oct 30, 2018
- Innova IGS 5, Innova IGS 6, Discovery IGS 7, Discovery IGS 7 ORK181403 · GE Healthcare · Nov 2, 2018
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- Oec 3DK233669 · Ge Oec Medical Systems, Inc. · Mar 28, 2024
- Remi Robotic Navigation SystemK223070 · Fusion Robotics, LLC · Oct 28, 2022
Recalls
Recalls FDA ties to K203346, newest first; each opens FDA's recall record| Mar 21, 2025 | Z-1820-2025OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field. | Open, Classified |
| Nov 21, 2024 | Z-0703-2025Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability. | Open, Classified |
| Jun 10, 2021 | Z-2160-2021There is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable. | Terminated |
Other OXO Clearances
The newest 510(k)s for the same product code; their median review was 152 days; see every OXO device- Digital C-ARM with 3DK253296 · Allengers Medical Systems Limited · Jun 5, 2026
- Orthoscan TAU MVP Mini C-Arm SystemK252579 · Ziehm-Orthoscan, Inc. · Jan 14, 2026
- OEC One CFDK253269 · Ge Hualun Medical Systems Co. , Ltd. · Nov 26, 2025
- Dual-Mode Mobile C-Arm (Geelin500A, Geelin500M)K251004 · Hefei Chimed Intelligent Machine Co., Ltd. · Nov 6, 2025
- Orthoscan TAU Mini C-ArmK250587 · Ziehm-Orthoscan, Inc. · Jul 2, 2025
- Orthoscan VERSA Mini C-ArmK243452 · Ziehm-Orthoscan, Inc. · Jan 14, 2025
- OEC One ASDK240828 · Ge Hualun Medical Systems Co. , Ltd. · Dec 27, 2024
- Oec 3DK233669 · Ge Oec Medical Systems, Inc. · Mar 28, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 13, 2020
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files