Arca
Skip to content
Tools
Tools

OEC Elite

Image-Intensified Fluoroscopic X-Ray System, Mobile. 510(k) K192819, from Ge Oec Medical Systems, Inc. (Salt Lake City, UT). Product code OXO, class 2.

As of Nov 20, 2026, K192819 is an FDA 510(k) held by Ge Oec Medical Systems, Inc., substantially equivalent on Nov 8, 2019 after 38 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 38 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • OEC EliteK172550 · Ge Oec Medical Systems, Inc. · Nov 16, 2017
  • OEC EliteK171565 · Ge Oec Medical Systems, Inc. · Aug 10, 2017

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K192819, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 21, 2025Z-1819-2025OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.Open
Nov 21, 2024Z-0704-2025Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.Open
May 15, 2023Z-2278-2023Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which…Open
Jun 10, 2021Z-2161-2021There is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.Terminated

Other OXO clearances

The newest 510(k)s for the same product code; their median review was 152 days; see every OXO device

Record

Type
Special
Received
Oct 1, 2019
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files