Interventional Fluoroscopic X-Ray System. 510(k) K171565, from Ge Oec Medical Systems, Inc. (Salt Lake City, UT). Product code OWB, class 2.

As of Nov 20, 2026, K171565 is an FDA 510(k) held by Ge Oec Medical Systems, Inc., substantially equivalent on Aug 10, 2017 after 72 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 72 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • OEC EliteK170752 · Oec Medical Systems, Inc. · May 19, 2017

Cited as a predicate

3 later 510(k) summaries name this device as a predicate
  • OEC One ASDK240828 · Ge Hualun Medical Systems Co. , Ltd. · Dec 27, 2024
  • OEC EliteK192819 · Ge Oec Medical Systems, Inc. · Nov 8, 2019
  • OEC EliteK172550 · Ge Oec Medical Systems, Inc. · Nov 16, 2017

Recalls

Recalls FDA ties to K171565, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 21, 2025Z-1819-2025OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.Open
Nov 21, 2024Z-0704-2025Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.Open
May 15, 2023Z-2278-2023Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which…Open
Jun 10, 2021Z-2161-2021There is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
May 30, 2017
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files