ARTIS pheno (VE2)

Interventional Fluoroscopic X-Ray System. 510(k) K201156, from Siemens Medical Solution USA, Inc. (Malvern, PA). Product code OWB, class 2.

As of Nov 18, 2026, K201156 is an FDA 510(k) held by Siemens Medical Solution USA, Inc., substantially equivalent on Jun 29, 2020 after 60 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 60 days
  • Recalls: 6
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K201156, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 10, 2022Z-1294-2022A video signal from one of the transceivers of the Large Display may be lost. In very rare cases, both transceivers of the Large Display could potentially fail which can lead to the loss of both video signals. This…Open
Feb 25, 2022Z-0987-2022Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block…Open
Feb 25, 2022Z-1339-2022After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient orientation/position may be misinterpreted and result in…Open
Oct 12, 2021Z-0515-2022Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be…Terminated
May 27, 2021Z-2029-2021Software error VE10B -In case of a fault within the drivetrain of an axis and while performing a stepped peripheral imaging (PERISTEPPING/PERIVISION), the C-arm may leave its intended travel path by up to 22 cm (9")…Open
Mar 19, 2021Z-1497-2021Siemens has become aware of a potential issue with software version VE20C. Planned procedures may have to be terminated and performed on an alternative x-ray system.Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Apr 30, 2020
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files