Arca
Skip to content
Tools
Tools

Monaco

System, Planning, Radiation Therapy Treatment. 510(k) K132971, from Impac Medical Systems, Inc. (Maryland Heights, MO). Product code MUJ, class 2.

As of Nov 19, 2026, K132971 is an FDA 510(k) held by Impac Medical Systems, Inc., substantially equivalent on Nov 20, 2013 after 58 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 58 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • MonacoK151233 · Impac Medical Systems, Inc. · Jul 7, 2015

Recalls

Recalls FDA ties to K132971, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 21, 2016Z-1009-2017Incorrect Enhanced Dynamic Wedge (EDW) or Virtual Wedge (VW) Calculations.Terminated
Nov 17, 2016Z-0668-2017Contours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm.Terminated
Dec 19, 2014Z-1237-2015Incorrect CT to ED when using the Monaco Image Statistics Tool.Terminated
Aug 18, 2014Z-2581-2014Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not…Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional
Received
Sep 23, 2013
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files