System, Planning, Radiation Therapy Treatment. 510(k) K151233, from Impac Medical Systems, Inc. (Maryland Heights, MO). Product code MUJ, class 2.

As of Nov 19, 2026, K151233 is an FDA 510(k) held by Impac Medical Systems, Inc., substantially equivalent on Jul 7, 2015 after 60 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 60 days
  • Recalls: 9

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K151233, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 28, 2023Z-1233-2023Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.Open
Jul 3, 2019Z-2091-2019When creating 3D plans using either MU or Dose weighting modes, if the user changes the number of fractions, rescales the plan and then changes the wedge angle, the Monitor Units are scaled incorrectly.Open
Jan 10, 2017Z-1044-2017Incorrect dose after editing beam number an wedge angle.Terminated
Dec 21, 2016Z-1009-2017Incorrect Enhanced Dynamic Wedge (EDW) or Virtual Wedge (VW) Calculations.Terminated
Nov 3, 2016Z-0660-2017Incorrect Dose when using the reset function.Terminated
Sep 29, 2016Z-0076-2017When creating 3D plans using either MU or Dose weighting modes, if the user changes the Physician's Intent Rx Dose and/or the number of fractions, and then modifies the wedge angle, the MU value is scaled incorrectly.Terminated
May 18, 2016Z-1879-2016When DICOM exporting a 3D Monaco plan and the "Composite Field Sequencing" (CFS) checkbox is selected, the Y jaws will snap to the port shape when they should remain where they were defined.Terminated
Oct 16, 2015Z-0181-2016Unintended update of Dose and MU and Incorrect Assignment of Bolus.Terminated
Sep 15, 2015Z-0112-2016Dose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities.Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional
Received
May 8, 2015
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files