Monaco Rtp System
System, Planning, Radiation Therapy Treatment. 510(k) K110730, from Computerized Medical Systems, Inc. (Maryland Heights, MO). Product code MUJ, class 2.
As of Nov 19, 2026, K110730 is an FDA 510(k) held by Computerized Medical Systems, Inc., substantially equivalent on Jun 24, 2011 after 100 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 100 days
- Recalls: 6
- Cited as a predicate by: 1 device
- Monaco Rtp System, Vmat OptionK091179 · Computerized Medical Systems, Inc. · Jul 16, 2009
- Ergo++ Version 1.7K080601 · 3D Line Medical Systems S.R.L. · Jun 6, 2008
- Eclipse Treatment Planning SystemK102011 · Varian Medical Systems, Inc. · Sep 3, 2010
- MonacoK132971 · Impac Medical Systems, Inc. · Nov 20, 2013
| Aug 18, 2016 | Z-2712-2016In a specific workflow where contours are edited (enlarged, moved, copied, etc.) and then the contour is deleted on some slices and saved in the same session, it is possible that the deleted contours are still present… | Terminated |
| Jul 2, 2013 | Z-2276-2013Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values. | Terminated |
| May 22, 2013 | Z-2277-2013Patient needs to be positioned as indicated by the treatment plan. | Terminated |
| Feb 11, 2013 | Z-0841-2013The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle. | Terminated |
| Jun 21, 2012 | Z-2007-2012Incorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports. | Terminated |
| Mar 5, 2012 | Z-1365-2012An unnecessary dose refresh can occur in certain situations which causes inconsistencies in the Beam Doses and the Total Dose. | Terminated |
- RT Elements (4.6)K261655 · Brainlab SE · Oct 5, 2026
- Virtual Cone (1.0)K260750 · Varian Medical Systems, Inc. · Jul 8, 2026
- RadOncAIK253050 · Informai · Jun 18, 2026
- Multimodality Simulation Workspace (MM Sim) (v1.0.0)K260207 · Philips Medical Systems Nederland B.V. · May 27, 2026
- eXaSkin PlusK252784 · Anatomical Geometry S.L. · May 22, 2026
- ART.1-USK253078 · Seetreat Pty, Ltd. · May 14, 2026
- syngo.via RT Image Suite VC10K252304 · Varian Medical Systems, Inc. · Mar 18, 2026
- Monaco RTP System (6.3)K252002 · Elekta Solutions AB · Feb 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 16, 2011
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files