Arca
Skip to content
Tools
Tools

Monaco Rtp System

System, Planning, Radiation Therapy Treatment. 510(k) K110730, from Computerized Medical Systems, Inc. (Maryland Heights, MO). Product code MUJ, class 2.

As of Nov 19, 2026, K110730 is an FDA 510(k) held by Computerized Medical Systems, Inc., substantially equivalent on Jun 24, 2011 after 100 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 100 days
  • Recalls: 6
  • Cited as a predicate by: 1 device

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • MonacoK132971 · Impac Medical Systems, Inc. · Nov 20, 2013

Recalls

Recalls FDA ties to K110730, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 18, 2016Z-2712-2016In a specific workflow where contours are edited (enlarged, moved, copied, etc.) and then the contour is deleted on some slices and saved in the same session, it is possible that the deleted contours are still present…Terminated
Jul 2, 2013Z-2276-2013Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values.Terminated
May 22, 2013Z-2277-2013Patient needs to be positioned as indicated by the treatment plan.Terminated
Feb 11, 2013Z-0841-2013The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.Terminated
Jun 21, 2012Z-2007-2012Incorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports.Terminated
Mar 5, 2012Z-1365-2012An unnecessary dose refresh can occur in certain situations which causes inconsistencies in the Beam Doses and the Total Dose.Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional
Received
Mar 16, 2011
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files