Eclipse Treatment Planning System
System, Planning, Radiation Therapy Treatment. 510(k) K102011, from Varian Medical Systems, Inc. (Palo Alto, CA). Product code MUJ, class 2.
As of Nov 19, 2026, K102011 is an FDA 510(k) held by Varian Medical Systems, Inc., substantially equivalent on Sep 3, 2010 after 49 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 49 days
- Recalls: 5
- Cited as a predicate by: 5 devices
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments.
- Modification To Eclipse Treatment Planning SystemK091492 · Varian Medical Systems, Inc. · Jun 18, 2009
- Radiance V2K133655 · Gmv Soluciones Globales Internet S.A.U. · Jan 31, 2014
- Smart Segmentation - Knowledge Based ContouringK133227 · Varian Medical Systems, Inc. · Mar 14, 2014
- MonacoK132971 · Impac Medical Systems, Inc. · Nov 20, 2013
- Eclipse Treatment Planning SystemK131891 · Varian Medical Systems, Inc. · Sep 25, 2013
- Monaco Rtp SystemK110730 · Computerized Medical Systems, Inc. · Jun 24, 2011
| Jan 28, 2013 | Z-0867-2013The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used. | Terminated |
| Aug 17, 2012 | Z-2462-2012Varian has identified an anomaly with the Eclipse Treatment Planning where an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not… | Terminated |
| Jan 30, 2012 | Z-1161-2012Varian has been identified an anomaly with the Eclipse Treatment Planning System where the calculated dose distribution is not updated or invalidated when a recalculation or reload is performed while the compensator or… | Terminated |
| Dec 12, 2011 | Z-0736-2012An anomaly has been identified with the Eclipse Treatment Planning System when planning for the Siemens Multileaf Collimator (MLC) where a virtual jaw can be positioned inside the MLC aperture and then used as the beam… | Terminated |
| May 20, 2011 | Z-2788-2011A software anomaly has been identified with the Eclipse Monte Carlo (eMC) algorithm where not-equidistant data points in the open beam profile used during configuration may lead to inaccurate results. | Terminated |
- RT Elements (4.6)K261655 · Brainlab SE · Oct 5, 2026
- Virtual Cone (1.0)K260750 · Varian Medical Systems, Inc. · Jul 8, 2026
- RadOncAIK253050 · Informai · Jun 18, 2026
- Multimodality Simulation Workspace (MM Sim) (v1.0.0)K260207 · Philips Medical Systems Nederland B.V. · May 27, 2026
- eXaSkin PlusK252784 · Anatomical Geometry S.L. · May 22, 2026
- ART.1-USK253078 · Seetreat Pty, Ltd. · May 14, 2026
- syngo.via RT Image Suite VC10K252304 · Varian Medical Systems, Inc. · Mar 18, 2026
- Monaco RTP System (6.3)K252002 · Elekta Solutions AB · Feb 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 16, 2010
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files