Oncentra, Oncentra External Beam, Oncentra Brachy

System, Planning, Radiation Therapy Treatment. 510(k) K121448, from Elekta Inc. D/B/A Nucletron Corporation (Baltimore, MD). Product code MUJ, class 2.

As of Nov 19, 2026, K121448 is an FDA 510(k) held by Elekta Inc. D/B/A Nucletron Corporation, substantially equivalent on Jul 2, 2012 after 48 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 48 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer. The treatment plans provide estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by qualified medical personnel.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate
  • MonacoK132971 · Impac Medical Systems, Inc. · Nov 20, 2013
  • Oncentra BrachyK132816 · Nucletron B.V. · Jun 13, 2014

Recalls

Recalls FDA ties to K121448, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 23, 2016Z-0987-2017Cross profile for Varian 60 degree wedge shows "horns."Terminated
Dec 18, 2015Z-0604-2016When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas…Terminated
May 21, 2015Z-1712-2015Incorrect dose calculation for Regions of Interest (ROIs) defined on a secondary image series.Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional
Received
May 15, 2012
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files