VERO Linear Accelerator System
Recall Z-2653-2014, part of FDA recall event 68909. See Mitsubishi Heavy Industries, Hiroshima Machinery Works's company page.
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK started this Class II device recall on Jul 5, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 1, 2015.
- Classification: Class II
- Started: Jul 5, 2014
- Status: Terminated
Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status. Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment
- Product
- VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
- Quantity
- 3 units in use in the US
- Lots and codes
- Software Version 3.0.0 and after. Serial Numbers 201902, 203901, and 203919
- Distribution
- Distributed in the states of NY, FL, and TX.
- Recalling firm
- MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK, HIROSHIMA, N/A, Japan
- How the firm notified
- FDA classified
- Sep 16, 2014
- VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator…Z-1308-2016 · Class II · Mar 7, 2016 · Terminated
- MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) Intended for radiation therapy of lesions,…Z-1244-2016 · Class II · Dec 15, 2015 · Terminated
- MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)Z-1574-2015 · Class II · Apr 17, 2015 · Terminated
- VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and…Z-0821-2015 · Class II · Dec 2, 2014 · Terminated
- MHI-TM2000 Linear Accelerator System, Software Version 3.0.0 revision 13 or earlier; Proprietary and Trade Name:…Z-1931-2013 · Class II · Jun 21, 2013 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports