Arca
Skip to content
Tools
Tools

MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)

Recall Z-1574-2015, part of FDA recall event 70638. See Mitsubishi Heavy Industries, Hiroshima Machinery Works's company page.

MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK started this Class II device recall on Apr 17, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 11, 2017.

  • Classification: Class II
  • Started: Apr 17, 2015
  • Status: Terminated

Why It Was Recalled

Class II

The operator console allows users to deliver therapeutic radiation to patients even though a specific communication error relevant to imaging conditions has occurred. In addition, the treatment record (the delivered radiation record) cannot be saved.

Record

Product
MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)
Quantity
11 units total
Lots and codes
Serial Numbers: 201902, 201903, 202903, 202905, 203901, 203902, 203903, 203905, 203906, 203919 & 203924
Distribution
Worldwide Distribution - US including FL, NY, OH, & TX, **Internationally - Belgium, Italy, Korea, France & Germany.
Recalling firm
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK, HIROSHIMA, N/A, Japan
How the firm notified
E-Mail
FDA classified
May 1, 2015

MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports