MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)
Recall Z-1574-2015, part of FDA recall event 70638. See Mitsubishi Heavy Industries, Hiroshima Machinery Works's company page.
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK started this Class II device recall on Apr 17, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 11, 2017.
- Classification: Class II
- Started: Apr 17, 2015
- Status: Terminated
The operator console allows users to deliver therapeutic radiation to patients even though a specific communication error relevant to imaging conditions has occurred. In addition, the treatment record (the delivered radiation record) cannot be saved.
- Product
- MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)
- Quantity
- 11 units total
- Lots and codes
- Serial Numbers: 201902, 201903, 202903, 202905, 203901, 203902, 203903, 203905, 203906, 203919 & 203924
- Distribution
- Worldwide Distribution - US including FL, NY, OH, & TX, **Internationally - Belgium, Italy, Korea, France & Germany.
- Recalling firm
- MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK, HIROSHIMA, N/A, Japan
- How the firm notified
- FDA classified
- May 1, 2015
- VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator…Z-1308-2016 · Class II · Mar 7, 2016 · Terminated
- MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) Intended for radiation therapy of lesions,…Z-1244-2016 · Class II · Dec 15, 2015 · Terminated
- VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and…Z-0821-2015 · Class II · Dec 2, 2014 · Terminated
- VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear…Z-2653-2014 · Class II · Jul 5, 2014 · Terminated
- MHI-TM2000 Linear Accelerator System, Software Version 3.0.0 revision 13 or earlier; Proprietary and Trade Name:…Z-1931-2013 · Class II · Jun 21, 2013 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports