VERO Linear Accelerator System
Recall Z-0821-2015, part of FDA recall event 69825. See Mitsubishi Heavy Industries, Hiroshima Machinery Works's company page.
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK started this Class II device recall on Dec 2, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Jun 26, 2017.
- Classification: Class II
- Started: Dec 2, 2014
- Status: Terminated
Software Anomaly. The VERO MHI-TM2000 Operator Console may provide Patient Positioning System (ExacTrac) with Image Angle Information Used for the 1st Port Image, for a subsequent port image fusion. As the result, ExacTrac may display a Digitally Reconstructed Radiograph (DRR) image taken from the angle for the 1st port image fusion, rather than the one taken from the angle for the intended port
- Product
- VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and 3.5.0.
- Quantity
- Domestic: 4 units
- Lots and codes
- Serial Numbers: 201902, 203901, 203919, and 203924 (US only)
- Distribution
- US Distribution to the states of NY, FL, TX and OH.
- Recalling firm
- MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK, HIROSHIMA, N/A, Japan
- How the firm notified
- FDA classified
- Dec 17, 2014
- VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator…Z-1308-2016 · Class II · Mar 7, 2016 · Terminated
- MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) Intended for radiation therapy of lesions,…Z-1244-2016 · Class II · Dec 15, 2015 · Terminated
- MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)Z-1574-2015 · Class II · Apr 17, 2015 · Terminated
- VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear…Z-2653-2014 · Class II · Jul 5, 2014 · Terminated
- MHI-TM2000 Linear Accelerator System, Software Version 3.0.0 revision 13 or earlier; Proprietary and Trade Name:…Z-1931-2013 · Class II · Jun 21, 2013 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports