VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator…
Recall Z-1308-2016, part of FDA recall event 73436. See Mitsubishi Heavy Industries, Hiroshima Machinery Works's company page.
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK started this Class II device recall on Mar 7, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 20, 2017.
- Classification: Class II
- Started: Mar 7, 2016
- Status: Terminated
In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is incorrectly set and if nobody is aware of it, an erroneous treatment would be given to the patient.
- Product
- VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
- Quantity
- 25 Worldwide, US: 4 units
- Lots and codes
- Software version 3.5.8, 3.5.2 and lower; Serial Numbers: 201902, 203901, 203919, 203924
- Distribution
- Nationwide Distribution including New York, Texas, Florida, and Ohio
- Recalling firm
- MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK, HIROSHIMA, N/A, Japan
- How the firm notified
- FDA classified
- Mar 31, 2016
- MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) Intended for radiation therapy of lesions,…Z-1244-2016 · Class II · Dec 15, 2015 · Terminated
- MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)Z-1574-2015 · Class II · Apr 17, 2015 · Terminated
- VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and…Z-0821-2015 · Class II · Dec 2, 2014 · Terminated
- VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear…Z-2653-2014 · Class II · Jul 5, 2014 · Terminated
- MHI-TM2000 Linear Accelerator System, Software Version 3.0.0 revision 13 or earlier; Proprietary and Trade Name:…Z-1931-2013 · Class II · Jun 21, 2013 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports