ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for:…

Recall Z-2426-2024, part of FDA recall event 94771. See Radiometer Medical's company page.

Radiometer Medical ApS started this Class II device recall on May 31, 2024 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: May 31, 2024
  • Status: Ongoing

Why It Was Recalled

Class II

Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.

Record

Product
ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for: " In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) " In vitro testing of samples of expired air for the parameters pO2 and pCO2 " In vitro testing of pleura samples for the pH parameter.
Quantity
27,127 analyzers
Lots and codes
Model Numbers: 393-800 and 393-801 Part Numbers/Serial Numbers/UDI(-DI) codes: ABL825 754R0100N003 754R0100N003 ABL825 754R0241N010 754R0241N010 ABL825 754R0334N006 754R0334N006 ABL825 754R0417N008 754R0417N008 ABL825 754R0449N004 754R0449N004 ABL825 754R0487N004 754R0487N004 ABL837 754R0631N006 754R0631N006 ABL825 754R0755N003 754R0755N003 ABL825 754R0757N007 754R0757N007 ABL825 754R0884N007 754R0884N007 ABL825 754R0917N002 754R0917N002 ABL825 754R0918N007 754R0918N007 ABL810 754R0941N008 754R0941N008 ABL827 754R1001N004 (01)05700693938004(10)754R1001N004 ABL805 754R1028N005 754R1028N005…
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Canada, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Dominican Republic, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Iran, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe Australia, New Zealand, New Caledonia, Tokelau, Austria, Belgium, Netherlands, China, Croatia, Czech Republic, Slovakia, Denmark, Iceland, Finland, Estonia, France, Germany, Hong Kong, Hungary, India, Brunei, Italy, Japan, Kazakhstan, Malaysia, Mexico, Norway, Poland, Russia, Singapore, South Africa, Cameroon, Ivory Coast, Malawi, South Korea, Spain, Portugal, Sweden, Switzerland, Turkey, United Kingdom, Ireland.
Recalling firm
Radiometer Medical ApS, Bronshoj, Denmark
How the firm notified
Letter
FDA classified
Jul 24, 2024

Radiometer Medical ApS's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.