Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90…

Recall Z-1659-2025, part of FDA recall event 96515. See Radiometer Medical's company page.

Radiometer Medical ApS started this Class II device recall on Mar 21, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Mar 21, 2025
  • Status: Ongoing

Why It Was Recalled

Class II

ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.

Record

Product
Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90 Series - Radiometer ABL90 FLEX PLUS Model Number 393-092 The Radiometer ABL90 Series is a portable, automated analyzer for use in the laboratory environment, near patients or point-of-care settings.
Quantity
6275 analyzers
Lots and codes
Model Number: 393-090 UDI-DI: 05700693930909 Model Number: 393-092 UDI-DI: 05700693930923 Serial Numbers: 090R0643N002 090R0749N001 090R0757N002 090R0757N004 090R0861N018 090R0874N009 090R0874N011 090R0874N012 090R0892N012 090R0897N023 090R0902N025 090R0943N010 090R0965N003 090R0965N007 090R0965N009 092R0150N001 092R0173N047 092R0173N049 092R0177N055 092R0180N011 092R0180N014 092R0180N025 092R0180N027 092R0185N022 092R0185N024 092R0185N025 092R0185N026 092R0187N016 092R0187N019 092R0187N021 092R0187N026 092R0187N027 092R0188N022 092R0188N024 092R0188N025 092R0188N026 092R0188N027 092R0188N028…
Distribution
U.S.: AL, AR, AZ, BM CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Canada, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Dominican Republic, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Iran, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe, Australia, New Zealand, New Caledonia, Tokelau, Austria, Belgium, Netherlands, China, Croatia, Czech Republic, Slovakia, Denmark, Iceland, Finland, Estonia, France, Germany, Hong Kong, Hungary, India, Brunei, Italy, Japan, Kazakhstan, Malaysia, Mexico, Norway, Poland, Russia, Singapore, South Africa, Cameroon, Ivory Coast, Malawi, South Korea, Spain, Portugal, Sweden, Switzerland, Turkey, United Kingdom, Ireland
Recalling firm
Radiometer Medical ApS, Bronshoj, N/A, Denmark
How the firm notified
Letter
FDA classified
Apr 23, 2025

Radiometer Medical ApS's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.