ABL800 Flex Analyzer Model Numbers 393-800 and 393-801

Recall Z-1321-2022, part of FDA recall event 90286. See Radiometer Medical's company page.

Radiometer Medical ApS started this Class II device recall on May 18, 2022 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: May 18, 2022
  • Status: Ongoing

Why It Was Recalled

Class II

There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.

Record

Product
ABL800 Flex Analyzer Model Numbers 393-800 and 393-801
Quantity
1489 analyzers
Lots and codes
All ABL800 Flex Analyzers containing software versions below V6.19 MR2; UDI/DI ABL810, 820, and 830, ABL805, ABL815, ABL825, ABL835: 05700693938004 ; UDI/DI ABL817, ABL827, ABL837: 05700693938011
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Bermuda, Canada, Italy, Norway, Austria, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe, China, Croatia, Denmark, Iceland, Germany, Spain, Portugal, India, Brunei, Japan, United Kingdom, Ireland, Russia, Hungary, Australia, New Zealand, Tokelau, Switzerland, France, Czech Republic, Slovakia, South Africa, Botswana, Namibia, Poland, Turkey, Sweden, Finland, Estonia, Kazakhstan, Malaysia, Hong Kong, Mexico, South Korea, Singapore.
Recalling firm
Radiometer Medical ApS, Bronshoj, N/A, Denmark
How the firm notified
Letter
FDA classified
Jul 1, 2022

Radiometer Medical ApS's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.