AQURE REF 933-599 Software Versions 2.5.2 2.5.3 2.5.4 2.6.0 2.6.1 The AQURE system is…

Recall Z-1394-2023, part of FDA recall event 91917. See Radiometer Medical's company page.

Radiometer Medical ApS started this Class II device recall on Mar 8, 2023 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Mar 8, 2023
  • Status: Ongoing

Why It Was Recalled

Class II

Due to potential software issue that may result in patient mix-up information.

Record

Product
AQURE REF 933-599 Software Versions 2.5.2 2.5.3 2.5.4 2.6.0 2.6.1 The AQURE system is intended to let you manage analytical devices and operator profiles. The user can associate patient data with test data. The system shows test results. The system receives data from connected devices at the point-of care or laboratory. It can send test results to the HIS/LIS. The system lets you send commands to selected devices. The system uses data related to the performance of devices, to tell users of issues to be managed. The AQURE system is intended for professional use.
Quantity
1,806 systems
Lots and codes
Model Number: 933-599 UDI-DI Code: 05700699335999 Software Versions: 2.5.2 and above (Versions 2.5.2, 2.5.3, 2.5.4, 2.60, and 2.6.1)
Distribution
Worldwide - U.S. Nationwide including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Brazil, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Paraguay, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay.
Recalling firm
Radiometer Medical ApS, Bronshoj, N/A, Denmark
How the firm notified
Letter
FDA classified
Apr 17, 2023

Radiometer Medical ApS's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.