lnvacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16. Packed one per carton. The…
Recall Z-2364-2016, part of FDA recall event 74581. See Invacare's company page.
Invacare Corporation started this Class II device recall on Jun 21, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Feb 7, 2017.
- Classification: Class II
- Started: Jun 21, 2016
- Status: Terminated
- Products in this event: 2
The rear gradation scale was screen printed upside down on the flowmeter gauge.
- Product
- lnvacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16. Packed one per carton. The lnvacare Precise Rx Pediatric Flowmeter adapts to lnvacare Concentrators for low-flow usage.The gauge has a scale printed on the front side and the back side. The front scale is the primary source for monitoring oxygen flow. The rear scale is an aide for monitoring oxygen flow.
- Quantity
- 395
- Lots and codes
- Serial numbers for Model IRCPF16: 16BF023724, 16BF023729, 16BF023734, 16BF023739, 16BF023744,16BF023720, 16BF023725, 16BF023730, 16BF023735, 16BF023740, 16BF023745,16BF023721, 16BF023726, 16BF023731, 16BF023736, 16BF023741, 16BF023746,16BF023722, 16BF023727, 16BF023732, 16BF023737, 16BF023742, 16BF023747,16BF023723, 16BF023728, 16BF023733, 16BF023738, 16BF023743, 16BF023748,16BF023749, 16BF030593, 16CF015205, 16CF019539, 16CF026064, 16CF030100,16BF023750, 16BF030594, 16CF015206, 16CF019540, 16CF026065, 16CF030101,16BF023751, 16CF016169, 16CF015207, 16CF019541, 16CF026066,…
- Distribution
- Worldwide Distribution - US to CA, CO, CT, FL, GA, IA, ID, IL,IN,KS,KY,LA,MA,MD,MI,MS,MT,NC,NE,NJ,NM,NY,OH,PA,SC, TN,TX,UT,VA,WA and Internationally to Canada, Germany and New Zealand.
- Recalling firm
- Invacare Corporation, Elyria, OH, United States
- How the firm notified
- Letter
- FDA classified
- Aug 8, 2016
- Invacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16 AW. Packed one per carton. The lnvacare Precise Rx"…Z-2365-2016 · Class II · Jun 21, 2016 · Terminated
Newest first, one product per recall event; search them all
- Invacare Homecare Series Bed and Invacare G-Series Beds Containing Component Model Numbers: (1) G50 (2) G53 (3)…Z-1251-2024 · Class II · Jan 22, 2024 · Ongoing
- DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): …Z-2445-2023 · Class II · Jun 15, 2023 · Ongoing
- Invacare Platinum 5NXG Oxygen Concentrator, Model Numbers P5NXG1, P5NXGC1 (Canada specific model)Z-1369-2022 · Class III · Jun 10, 2022 · Ongoing
- Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2VZ-1130-2022 · Class II · Mar 30, 2022 · Ongoing
- Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and…Z-0447-2022 · Class II · Nov 12, 2021 · Ongoing
- Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental…Z-0514-2021 · Class II · Oct 30, 2020 · Terminated
- Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered…Z-0549-2019 · Class II · Nov 16, 2018 · Terminated
- Assist Bed Rail - 6632Z-2218-2017 · Class II · Apr 25, 2017 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports