Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory…

Recall Z-0514-2021, part of FDA recall event 86604. See Invacare's company page.

Invacare Corporation started this Class II device recall on Oct 30, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 31, 2024.

  • Classification: Class II
  • Started: Oct 30, 2020
  • Status: Terminated

Why It Was Recalled

Class II

Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.

Record

Product
Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC5PO2V (US), IRC5PO2VC (Canada), All World: IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL4, IRC5PO2VAWL5
Quantity
384,767 units
Lots and codes
Model numbers: IRC5PO2V from Serial # 14HF039890 to 18JF007415; IRC5PO2VC from Serial # 14IF004028 to 18JF028633; IRC5PO2VAW from serial #17HF030338-18IF018522; IRC5PO2VAWL2 from serial # 18BF007341-18IF024451; IRC5PO2VAWL3 from serial # 17JF002563-19AF008559; IRC5PO2VAWL5 from serial # 18EF003466-18EF003544. DI Number: 00841447100249
Distribution
Worldwide distribution. US Nationwide, Canada, Australia, China, Germany, Guam, New Zealand, and Northern Mariana Islands
Recalling firm
Invacare Corporation, Elyria, OH, United States
How the firm notified
Letter
FDA classified
Dec 1, 2020

Invacare Corporation's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.