Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory…
Recall Z-0514-2021, part of FDA recall event 86604. See Invacare's company page.
Invacare Corporation started this Class II device recall on Oct 30, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 31, 2024.
- Classification: Class II
- Started: Oct 30, 2020
- Status: Terminated
Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.
- Product
- Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC5PO2V (US), IRC5PO2VC (Canada), All World: IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL4, IRC5PO2VAWL5
- Quantity
- 384,767 units
- Lots and codes
- Model numbers: IRC5PO2V from Serial # 14HF039890 to 18JF007415; IRC5PO2VC from Serial # 14IF004028 to 18JF028633; IRC5PO2VAW from serial #17HF030338-18IF018522; IRC5PO2VAWL2 from serial # 18BF007341-18IF024451; IRC5PO2VAWL3 from serial # 17JF002563-19AF008559; IRC5PO2VAWL5 from serial # 18EF003466-18EF003544. DI Number: 00841447100249
- Distribution
- Worldwide distribution. US Nationwide, Canada, Australia, China, Germany, Guam, New Zealand, and Northern Mariana Islands
- Recalling firm
- Invacare Corporation, Elyria, OH, United States
- How the firm notified
- Letter
- FDA classified
- Dec 1, 2020
Newest first, one product per recall event; search them all
- Invacare Homecare Series Bed and Invacare G-Series Beds Containing Component Model Numbers: (1) G50 (2) G53 (3)…Z-1251-2024 · Class II · Jan 22, 2024 · Ongoing
- DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): …Z-2445-2023 · Class II · Jun 15, 2023 · Ongoing
- Invacare Platinum 5NXG Oxygen Concentrator, Model Numbers P5NXG1, P5NXGC1 (Canada specific model)Z-1369-2022 · Class III · Jun 10, 2022 · Ongoing
- Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2VZ-1130-2022 · Class II · Mar 30, 2022 · Ongoing
- Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and…Z-0447-2022 · Class II · Nov 12, 2021 · Ongoing
- Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered…Z-0549-2019 · Class II · Nov 16, 2018 · Terminated
- Assist Bed Rail - 6632Z-2218-2017 · Class II · Apr 25, 2017 · Terminated
- Junction Box - 1104343 and 1193560Z-1185-2017 · Class II · Dec 2, 2016 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports