Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2V
Recall Z-1130-2022, part of FDA recall event 90084. See Invacare's company page.
Invacare Corporation started this Class II device recall on Mar 30, 2022 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Mar 30, 2022
- Status: Ongoing
The battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to the wheelchair. The disconnection of the battery harness results in loss of power to the wheelchair.
- Product
- Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2V
- Quantity
- 133 devices
- Lots and codes
- UDI (DI): Model Number Device Identifier TDXSP2 00841447100911 TDXSP2-MCG 00841447100966 TDXSP2V 00841447101512 Serial Numbers: 22CE000031 22BE004288 22BE002669 22BE003672 22BE004782 22BE004151 22CE000984 22BE004286 22BE004171 22BE004540 22CE001086 22BE003661 22CE000184 22CE000878 22CE001449 22CE000032 22CE001421 22BE004933 22CE001002 22CE000180 22CE000725 22BE004544 22BE005009 22BE003671 22BE004284 22BE004159 22BE004285 22BE002652 22BE004282 22BE004798 22BE004281 22BE004170 22CE000182 22CE000183 22CE000876 22CE000873 22CE000981 22CE000870 22CE000871 22CE001248 22CE001004 22CE001084…
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AR, CA, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA, and WI. The country of Canada.
- Recalling firm
- Invacare Corporation, Elyria, OH, United States
- How the firm notified
- Letter
- FDA classified
- May 25, 2022
Newest first, one product per recall event; search them all
- Invacare Homecare Series Bed and Invacare G-Series Beds Containing Component Model Numbers: (1) G50 (2) G53 (3)…Z-1251-2024 · Class II · Jan 22, 2024 · Ongoing
- DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): …Z-2445-2023 · Class II · Jun 15, 2023 · Ongoing
- Invacare Platinum 5NXG Oxygen Concentrator, Model Numbers P5NXG1, P5NXGC1 (Canada specific model)Z-1369-2022 · Class III · Jun 10, 2022 · Ongoing
- Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and…Z-0447-2022 · Class II · Nov 12, 2021 · Ongoing
- Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental…Z-0514-2021 · Class II · Oct 30, 2020 · Terminated
- Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered…Z-0549-2019 · Class II · Nov 16, 2018 · Terminated
- Assist Bed Rail - 6632Z-2218-2017 · Class II · Apr 25, 2017 · Terminated
- Junction Box - 1104343 and 1193560Z-1185-2017 · Class II · Dec 2, 2016 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports