Arca
Skip to content
Tools
Tools

Invacare 9805 and 9805P Hydraulic Patient Lifts

Recall Z-0549-2019, part of FDA recall event 81071. See Invacare's company page.

Invacare Corporation started this Class II device recall on Nov 16, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 15, 2020.

  • Classification: Class II
  • Started: Nov 16, 2018
  • Status: Terminated

Why It Was Recalled

Class II

Invacare has identified via customer complaints, the potential for the mounting bolt connecting the caster to the base frame of the patient lifts to become loose. Loose hardware can cause wear to the hardware and housings. If left unresolved, the caster may separate from the base of the lift causing the Lift to tilt and a patient fall. Failure does not occur right away but happens over time if the lift is not properly maintained. Patient falls are inherently dangerous and may result in a range of injuries that are dependent on physical condition of the patient and the nature of the fall. When injuries occur, they are typically non-serious injuries such as bruising, lacerations, or abrasions. In rare cases, more serious injuries such as bone fracture, head trauma, or death can occur. For caregivers using the lift, attempts to manually stabilize a lift if it becomes unstable could result in sprains, serious back injuries, or bone fractures.

Record

Product
Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered patient lift is to lift and transfer a patient from one surface to another, as from a bed to a wheelchair. The device includes straps and a sling to support the patient.
Quantity
73,703 units
Lots and codes
Serial numbers ranging between 09FSZ590001 to 16ESZ104075. Specifically, serial numbers with the following starting values (see below). Each consignee will be receiving an impacted units list, which will include full serial numbers for the units that they have purchased: 10ASZ 11ASZ 12ASZ 13ASZ 14ASZ 15ASZ 16ASZ 10BSZ 11BSZ 12BSZ 13BSZ 14BSZ 15BSZ 16BSZ 10CSZ 11CSZ 12CSZ 13CSZ 14CSZ 15CSZ 16CSZ 10DSZ 11DSZ 12DSZ 13DSZ 14DSZ 15DSZ 16DSZ 10ESSZ 11ESSZ 12ESSZ 13ESSZ 14ESSZ 15ESSZ 16ESSZ 09FSZ 10FSZ 11FSZ 12FSZ 13FSZ 14FSZ 15FSZ 09GSZ 10GSZ 11GSZ 12GSZ 13GSZ 14GSZ 15GSZ 09HSZ 10HSZ 11HSZ 12HSZ…
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Canada, Colombia, Germany, Guam, India, Northern Mariana Islands, Philippines and South Africa.
Recalling firm
Invacare Corporation, Elyria, OH, United States
How the firm notified
Letter
FDA classified
Dec 3, 2018

Invacare Corporation's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.