OEC Elite REF Image-Intensified Fluoroscopic X-Ray System
Recall Z-0704-2025, part of FDA recall event 95846. See GE OEC Medical Systems's company page.
GE OEC Medical Systems, Inc started this Class II device recall on Nov 21, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Nov 21, 2024
- Status: Ongoing
- Products in this event: 4
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
- Product
- OEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.
- Quantity
- 100 systems
- Lots and codes
- Product Name: OEC Elite GTIN code: 00001234567895 Lot/Serial Numbers / UDI codes: FAXXXE00254 (01)00840682121729(11)190400(21)FAXXXE00254 FBHXTX00837 (01)00840682121729(11)230300(21)FBHXTX00837 FBHXTX00912 (01)00840682121729(11)230500(21)FBHXTX00912 FBHXTX01278 (01)00840682121729(11)240800(21)FBHXTX01278 FBXXXE00731 (01)00840682121729(11)190500(21)FBXXXE00731 FBMHTX00470 (01)00840682121729(11)240900(21)FBMHTX00470 F2XXTX00801 (01)00840682121729(11)240800(21)F2XXTX00801 FBHXXX00171 (01)00840682121729(11)180600(21)FBHXXX00171 FBXXXE00715 (01)00840682121729(11)190500(21)FBXXXE00715 FBMHTX00444…
- Distribution
- U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
- Recalling firm
- GE OEC Medical Systems, Inc, Salt Lake City, UT, United States
- How the firm notified
- Letter
- FDA classified
- Dec 18, 2024
- OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging…Z-0703-2025 · Class II · Nov 21, 2024 · Ongoing
- OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy system is…Z-0705-2025 · Class II · Nov 21, 2024 · Ongoing
- OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide…Z-0706-2025 · Class II · Nov 21, 2024 · Ongoing
Newest first, one product per recall event; search them all
- OEC Elite Systems with 9-inch Image IntensifierZ-2278-2023 · Class II · May 15, 2023 · Ongoing
- OEC Elite mobile C-arm - Product Usage: intended to provide fluoroscopic and digital spot images of the patient…Z-2161-2021 · Class II · Jun 10, 2021 · Terminated
- OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical SystemsZ-1469-2017 · Class II · Feb 21, 2017 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports