Arca
Skip to content
Tools
Tools

OEC 3D REF Image-Intensified Fluoroscopic X-Ray System

Recall Z-0703-2025, part of FDA recall event 95846. See GE OEC Medical Systems's company page.

GE OEC Medical Systems, Inc started this Class II device recall on Nov 21, 2024 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Nov 21, 2024
  • Status: Ongoing
  • Products in this event: 4

Why It Was Recalled

Class II

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Record

Product
OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.
Quantity
84 systems
Lots and codes
Product Name: OEC 3D GTIN code: 00195278096944 Lot/Serial Numbers / UDI codes: FBMHDX00436 (01)00195278096944(11)240800(21)FBMHDX00436 FBMHDX00352 (01)00195278096944(11)240500(21)FBMHDX00352 FBMHDX00455 (01)00195278096944(11)240900(21)FBMHDX00455 FBMHDX00384 (01)00195278096944(11)240600(21)FBMHDX00384 FBMHDX00434 (01)00195278096944(11)240900(21)FBMHDX00434 FBMHDX00443 (01)00195278096944(11)240900(21)FBMHDX00443 FBMHDX00407 (01)00195278096944(11)240700(21)FBMHDX00407 FBMHDX00363 (01)00195278096944(11)240500(21)FBMHDX00363 FBMHDX00424 (01)00195278096944(11)240800(21)FBMHDX00424 FBMHDX00442…
Distribution
U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
Recalling firm
GE OEC Medical Systems, Inc, Salt Lake City, UT, United States
How the firm notified
Letter
FDA classified
Dec 18, 2024

Other Products in This Recall

Recall event 95846

GE OEC Medical Systems, Inc's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports