Arca
Skip to content
Tools
Tools

OEC Elite mobile C-arm - Product Usage: intended to provide fluoroscopic and digital spot…

Recall Z-2161-2021, part of FDA recall event 88202. See GE OEC Medical Systems's company page.

GE OEC Medical Systems, Inc started this Class II device recall on Jun 10, 2021 (voluntary: firm initiated); FDA lists it as terminated, terminated May 9, 2024.

  • Classification: Class II
  • Started: Jun 10, 2021
  • Status: Terminated
  • Products in this event: 2

Why It Was Recalled

Class II

There is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.

Record

Product
OEC Elite mobile C-arm - Product Usage: intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.
Quantity
3570 units
Lots and codes
All OEC Elite systems manufactured on or after January 2019
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands and the countries of Colombia, Egypt, Peru, Chile, Spain, France, Germany, Hong Kong, Canada, Japan, Brazil, Turkey, United Arab Emirates, Singapore, Australia, Thailand, United Kingdom, Italy, Korea, Russia, CHINA, Romania, Nigeria, Czechia, Argentina, Taiwan, Bangladeh, Malaysia, Bosnia and Herzegovina, Poland, Belgium, Indonesia, Dominican Republic, Switzerland, Hungary, New Zealand, Vietnam, Mexico, India, Norway, Philippines, Ecuador, Lithuania, HONDURAS, Uruguay, Sweden, Croatia, Jamaica, Kuwait, Ethiopia, Saudi Arabia, Finland, Bahamas, Trinidad and Tobago, Pakistan, Slovakia, Costa Rica, Israel, Panama, Greece, South Africa and DENMARK.
Recalling firm
GE OEC Medical Systems, Inc, Salt Lake City, UT, United States
How the firm notified
Letter
FDA classified
Jul 29, 2021

Other Products in This Recall

Recall event 88202

GE OEC Medical Systems, Inc's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports