OEC Elite Systems with 9-inch Image Intensifier
Recall Z-2278-2023, part of FDA recall event 92504. See GE OEC Medical Systems's company page.
GE OEC Medical Systems, Inc started this Class II device recall on May 15, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: May 15, 2023
- Status: Ongoing
- Products in this event: 5
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
- Product
- OEC Elite Systems with 9-inch Image Intensifier
- Quantity
- 812
- Lots and codes
- UDI-DI: 00840682121729 Serial Numbers: FSXXTX00180, FSXXXX00020, F9XXTX00007, F9XXTX00008, F9XXXX00021, FSXXTX00068, FSXXTX00303, FSXXXX00054, FSXXTX00333, FSXXTX00272, F9XXXX00046, FSXXXX00009, FSXXXX00005, FSXXTX00350, F9XXTX00025, FSXXTX00131, FSXXTX00103, FSXXTX00116, FSXXTX00073, FSXXXX00056, FSXXTX00120, FSXXTX00171, FSXXTX00307, FSXXXX00006, FSXXTX00153, FSXXTX00154, FSXXTX00214, FSXXTX00048, FSXXTX00062, FSXXTX00170, FSXXTX00172, FSXXTX00362, FSXXTX00368, FSXXXX00024, F9XXTX00202, FSXXTX00326, FSXXTX00260, FSXXTX00133, FSXXXX00011, FSXXTX00286, F9XXTX00045, FSXXTX00241, FSXXTX00179,…
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
- Recalling firm
- GE OEC Medical Systems, Inc, Salt Lake City, UT, United States
- How the firm notified
- Letter
- FDA classified
- Jul 31, 2023
- OEC Flexiview 8800 Systems with 9-inch Image IntensifierZ-2275-2023 · Class II · May 15, 2023 · Ongoing
- Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N…Z-2279-2023 · Class II · May 15, 2023 · Ongoing
- OEC 9900 Systems with 9-inch Image IntensifierZ-2277-2023 · Class II · May 15, 2023 · Ongoing
- OEC 9800 Systems with 9-inch Image IntensifierZ-2276-2023 · Class II · May 15, 2023 · Ongoing
Newest first, one product per recall event; search them all
- OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide…Z-0706-2025 · Class II · Nov 21, 2024 · Ongoing
- OEC Elite mobile C-arm - Product Usage: intended to provide fluoroscopic and digital spot images of the patient…Z-2161-2021 · Class II · Jun 10, 2021 · Terminated
- OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical SystemsZ-1469-2017 · Class II · Feb 21, 2017 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports