Invacare AVIVA FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers…
Recall Z-0446-2022, part of FDA recall event 89062. See Invacare's company page.
Invacare Corporation started this Class II device recall on Nov 12, 2021 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Nov 12, 2021
- Status: Ongoing
- Products in this event: 3
Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury
- Product
- Invacare AVIVA FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: IFX-20R, IFX-20SP, IFX-20MP, IFX-20C
- Quantity
- 303 units US; 20 units OUS
- Lots and codes
- LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: IFX-20C 00841447111627 IFX-20MP 00841447111658 IFX-20R 00841447111634 IFX-20SP 00841447111641 IFX20WIDE 00841447111658
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.
- Recalling firm
- Invacare Corporation, Elyria, OH, United States
- How the firm notified
- Letter
- FDA classified
- Jan 3, 2022
- Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) …Z-0445-2022 · Class II · Nov 12, 2021 · Ongoing
- Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and…Z-0447-2022 · Class II · Nov 12, 2021 · Ongoing
Newest first, one product per recall event; search them all
- Invacare Homecare Series Bed and Invacare G-Series Beds Containing Component Model Numbers: (1) G50 (2) G53 (3)…Z-1251-2024 · Class II · Jan 22, 2024 · Ongoing
- DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): …Z-2445-2023 · Class II · Jun 15, 2023 · Ongoing
- Invacare Platinum 5NXG Oxygen Concentrator, Model Numbers P5NXG1, P5NXGC1 (Canada specific model)Z-1369-2022 · Class III · Jun 10, 2022 · Ongoing
- Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2VZ-1130-2022 · Class II · Mar 30, 2022 · Ongoing
- Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental…Z-0514-2021 · Class II · Oct 30, 2020 · Terminated
- Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered…Z-0549-2019 · Class II · Nov 16, 2018 · Terminated
- Assist Bed Rail - 6632Z-2218-2017 · Class II · Apr 25, 2017 · Terminated
- Junction Box - 1104343 and 1193560Z-1185-2017 · Class II · Dec 2, 2016 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports