Bayer Advia 1650 Chemistry System
Electrode, Ion Specific, Sodium. 510(k) K990346, from Bayer Corp. (Tarrytown, NY). Product code JGS, class 2.
As of Nov 17, 2026, K990346 is an FDA 510(k) held by Bayer Corp., substantially equivalent on May 21, 1999 after 106 days, with 14 recalls tied to it in FDA's recall database.
- Substantially equivalent: May 21, 1999
- Review time: 106 days
- Recalls: 14
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe ADVIA 1650 Chemistry System is an automated, clinical chemistry analyzer that can run tests on human serum, plasma, or urine in random access, batch, and STAT modes at a throughput rate of 1200 photometric tests per hour and 450 electrolyte tests per hour. The photometric analyzer performs clinical chemistry and immuno-turbidimetric methods.
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- VITROS Chemistry Products Cl- SlidesK162020 · Ortho Clinical Diagnostics · Dec 16, 2016
- Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A,…K151767 · Siemens Healthcare Diagnostics, Inc. · Dec 11, 2015
- Vitros 5,1 Fs Chemistry System, Vitros Chemistry Products Dhdl Reagent Pack, Calibrator…K031924 · Ortho-Clinical Diagnostics · Aug 7, 2003
Recalls
Recalls FDA ties to K990346, newest first; each opens FDA's recall record| May 18, 2021 | Z-2014-2021ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously… | Terminated |
| Oct 7, 2015 | Z-0755-2016Multiple Software issues. Auto Start-UP Fail, Calibration Interval Resets when a Reagent Blank is run, Control Definition screen assumes range defined is 2 SD, Laboratory Information System (LIS)… | Terminated |
| Aug 28, 2015 | Z-0183-2016ISE module may produce discrepant, unflagged results for Sodium, Potassium or Chloride when the respective electrode is left unplugged after replacement, washing or maintenance activities. The calibration and 2-level QC… | Terminated |
| Feb 11, 2015 | Z-1399-2015Siemens Healthcare Diagnostics has confirmed an issue with ADVIA Chemistry XPT software version 1.0.2. The system may cause samples to remain in an Inprocess state. Test results on a sample that is held Inprocess will… | Terminated |
| Dec 17, 2013 | Z-0800-2014An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits,… | Terminated |
| Nov 6, 2013 | Z-0940-2014The ADVIA Chemistry software systems using certain software versions allows the system to automatically run assay calibrations and quality control (QC). | Terminated |
| Sep 3, 2013 | Z-0344-2014Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or 2400 Chemistry Systems. | Terminated |
| Sep 3, 2013 | Z-0343-2014Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or 2400 Chemistry Systems. | Terminated |
| Sep 8, 2006 | Z-0353-2007Bayer received a customer complaint where the Valproic Acid Quality control material recovered lower than target when it was run in a panel of other Therapeutic Drug (TDM) assays. The under recovery observed by the… | Terminated |
| Sep 8, 2006 | Z-0352-2007Bayer received a customer complaint where the Valproic Acid Quality control material recovered lower than target when it was run in a panel of other Therapeutic Drug (TDM) assays. The under recovery observed by the… | Terminated |
| Sep 8, 2006 | Z-0351-2007Bayer received a customer complaint where the Valproic Acid Quality control material recovered lower than target when it was run in a panel of other Therapeutic Drug (TDM) assays. The under recovery observed by the… | Terminated |
| Feb 17, 2006 | Z-1448-06Low QC recovery observed on carbamazepine (CARB) results immediately following a Gentamicin (GENT) assay or a Digoxin (DIG) assay. When GENT or DIG precedes CARB, the CARB test result is artificially low (as much as… | Terminated |
| Feb 17, 2006 | Z-1450-06Low QC recovery observed on carbamazepine (CARB) results immediately following a Gentamicin (GENT) assay or a Digoxin (DIG) assay. When GENT or DIG precedes CARB, the CARB test result is artificially low (as much as… | Terminated |
| Feb 17, 2006 | Z-1449-06Low QC recovery observed on carbamazepine (CARB) results immediately following a Gentamicin (GENT) assay or a Digoxin (DIG) assay. When GENT or DIG precedes CARB, the CARB test result is artificially low (as much as… | Terminated |
Other JGS Clearances
The newest 510(k)s for the same product code; their median review was 268 days; see every JGS device- ABL90 FLEX PLUS SystemK252475 · Radiometer Medicals Aps · May 1, 2026
- i-STAT CG8+ cartridge with the i-STAT 1 SystemK230275 · Abbott Point of Care, Inc. · Oct 27, 2023
- A-LYTE® Integrated Multisensor (IMT Na K Cl)K222438 · Siemens Healthcare Diagnostics, Inc. · Jul 25, 2023
- ISE Reagents, Glucose, CRP Latex, DxC 500 AU Clinical Chemistry AnalyzerK220977 · Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd. · Jul 20, 2023
- GLP systems TrackK213486 · Abbott Laboratories · Mar 10, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 4, 1999
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1665
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files