Vitros 5,1 FS Chemistry System, Vitros Chemistry Products Dhdl Reagent Pack, Calibrator Kit 19, FS Calibrator 1
Calibrator, Secondary. 510(k) K031924, from Ortho-Clinical Diagnostics (Rochester, NY). Product code JIT, class 2.
As of Nov 21, 2026, K031924 is an FDA 510(k) held by Ortho-Clinical Diagnostics, substantially equivalent on Aug 7, 2003 after 45 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 45 days
- Recalls: 25+
- Bayer Advia 1650 Chemistry SystemK990346 · Bayer Corp. · May 21, 1999
- Direct Hdl-Cholestrol LiquidK982341 · Randox Laboratories, Ltd. · Jul 30, 1998
| May 9, 2025 | Z-2040-2025a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date… | Open |
| Jan 26, 2024 | Z-1230-2024A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules,… | Open |
| Jan 26, 2024 | Z-1225-2024A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported. | Open |
| Oct 10, 2023 | Z-0544-2024During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen… | Open |
| Aug 9, 2023 | Z-0008-2024Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an… | Open |
| Dec 21, 2015 | Z-1073-2016, Z-1075-2016Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or… | Terminated |
| Nov 3, 2015 | Z-0642-2016, Z-0643-2016Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the… | Terminated |
| Sep 10, 2015 | Z-0231-2016Software Anomaly during ADD Installation on VITROS 5,1 FS Chemistry Systems using Software Versions 2.2.1 through 2.8. The user modified parameters would not be saved as expected, but remained changed to the default… | Terminated |
| Jun 30, 2015 | Z-2544-2015Software anomaly may occur during the installation of an Assay Data Diskette (ADD) using Software Versions 2.2.1 through 2.8. If the user is utilizing User- Configured (Modified) parameters, ALL default settings are… | Terminated |
| Mar 10, 2015 | Z-1521-2015Software Anomaly; It is possible for the device to process samples with a cartridge other than the intended cartridge, potentially leading to erroneous patient results. If this anomaly occurred previously, an indication… | Terminated |
| Nov 19, 2014 | Z-0969-2015, Z-0968-2015, Z-0967-2015Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration. | Terminated |
| Sep 29, 2014 | Z-1004-2015Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2. | Terminated |
| Sep 8, 2014 | Z-0077-2015Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator… | Terminated |
| Jul 7, 2014 | Z-2140-2014Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with… | Terminated |
| Apr 10, 2014 | Z-1734-2014, Z-1736-2014Potential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS 4600 Chemistry System, the VITROS 5600 Integrated System and the VITROS 5,1 Chemistry System. MW codes may… | Terminated |
| Jun 6, 2013 | Z-1948-2013Ortho Clinical Diagnostics is recalling CM/RT wear pads which are a component in the CM/RT incubator due to becoming worn before their replacement. | Terminated |
| May 6, 2013 | Z-1647-2013, Z-1648-2013There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt. | Terminated |
| Feb 7, 2013 | Z-1107-2013, Z-1106-2013Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an… | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 23, 2003
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1150
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files